ResMed S9 VPAP Tx manual Flow maximum at set pressures, Displayed values, Min, Value Accuracy

Page 33

Flow (maximum) at set pressures

The following are measured at the end of the specified air tubing:

Pressure

S9 VPAP Tx and

S9 VPAP Tx, H5i

S9 VPAP Tx and

S9 VPAP Tx, H5i

cm H2O

standard air

and standard air

SlimLine,

and ClimateLine,

 

tubing,

tubing,

L/min

L/min

 

L/min

L/min

 

 

4

200

170

195

170

8

200

170

190

170

12

200

170

184

170

16

200

170

175

170

20

190

170

168

161

25

180

161

144

125

Displayed values

Value

Range

Display resolution

Pressure sensor at air outlet

Mask pressure

3–30 cm H2O

0.1 cm H2O

Flow derived values

 

 

Leak

0–200 L/min

1 L/min

Tidal volume

0–4000 mL

1 mL

Respiratory rate

0–50 BPM

1 BPM

Minute ventilation

0–30 L/min

0.1 L/min

Ti

0.1–4.0 sec

0.1 sec

I:E ratio

1:50–2:1

0.1

Value

Accuracy1

 

Pressure measurement1

 

 

Mask pressure

±0.5 cm H2O (+ 4% of measured value)

Flow measurements1

 

 

Leak2

±12 L/min or 20% of reading, whichever is greater, at 0 to 60 L/min

Tidal volume2

±20%

 

Respiratory rate2

±1.0 BPM

 

Minute ventilation2

±20%

 

1Results are expressed at ATPD (Ambient Temperature and Pressure, Dry)

2Accuracy may be reduced by the presence of leaks, supplemental oxygen, tidal volumes <100 mL or minute ventilation <3 L/min.

29 English

Image 33
Contents S9 Vpap Tx Page S9 Vpap Tx Contents General warnings and cautions S9 Vpap Tx indications for use WelcomeS9 Vpap Tx contraindications S9 Vpap Tx adverse effectsS9 Vpap Tx at a glance Supplemental oxygen MasksAntibacterial filters HumidifierTherapy modes Therapy mode DescriptionKey Function S9 Vpap Tx basicsSetting up the S9 Vpap Tx Filling the water tub English Mask and air tubing settings Mask type SettingsAir tubing Specifications Settings Navigating the menus About the menusHome menu Humidity level Changing settings via the Home menuClimate Control Mask FitViewing the treatment screens Parameter Modes Description Treatment screen parametersSetup menu access Setup menuSettings ConfigurationParameter Modes Clinical setup menu parametersRange Description Ü ü ü ü ü ü ü üÜ ü ü Parameter Modes Range DescriptionLED Service Info menuParameter Description Managing Climate Control TemperatureMode Heated tubeDelivering therapy To add supplemental oxygen Adding supplemental oxygenFilter Efficiency Cleaning and maintenanceWorking with alarms Alarms menuTesting the alarms Problem/Possible cause Solution TroubleshootingNo display Insufficient air delivered from the device Water splashing on your face from the H5i Alarm is activated and the LCD screen display disappears Leaking water tub Air feels too warm/cold in the maskTechnical specifications General technical specificationsS9 Vpap Tx technical specifications Pneumatic flow path Flow maximum at set pressures Value Range Display resolutionDisplayed values MinServicing SymbolsProduct Limited warrantyPage GLO Multi