Flow (maximum) at set pressures
The following are measured at the end of the specified air tubing:
Pressure | S9 VPAP Tx and | S9 VPAP Tx, H5i | S9 VPAP Tx and | S9 VPAP Tx, H5i |
cm H2O | standard air | and standard air | SlimLine, | and ClimateLine, |
| tubing, | tubing, | L/min | L/min |
| L/min | L/min |
|
|
4 | 200 | 170 | 195 | 170 |
8 | 200 | 170 | 190 | 170 |
12 | 200 | 170 | 184 | 170 |
16 | 200 | 170 | 175 | 170 |
20 | 190 | 170 | 168 | 161 |
25 | 180 | 161 | 144 | 125 |
Displayed values
Value
Range
Display resolution
Pressure sensor at air outlet
Mask pressure | 0.1 cm H2O | |
Flow derived values |
|
|
Leak | 1 L/min | |
Tidal volume | 1 mL | |
Respiratory rate | 1 BPM | |
Minute ventilation | 0.1 L/min | |
Ti | 0.1 sec | |
I:E ratio | 0.1 | |
Value | Accuracy1 |
|
Pressure measurement1 |
|
|
Mask pressure | ±0.5 cm H2O (+ 4% of measured value) | |
Flow measurements1 |
|
|
Leak2 | ±12 L/min or 20% of reading, whichever is greater, at 0 to 60 L/min | |
Tidal volume2 | ±20% |
|
Respiratory rate2 | ±1.0 BPM |
|
Minute ventilation2 | ±20% |
|
1Results are expressed at ATPD (Ambient Temperature and Pressure, Dry)
2Accuracy may be reduced by the presence of leaks, supplemental oxygen, tidal volumes <100 mL or minute ventilation <3 L/min.