ResMed S9 VPAP Tx manual Clinical setup menu parameters, Parameter Modes, Range Description, Ü ü

Page 20

Clinical setup menu parameters*

Parameter

Modes

CPAP

AutoSet

VAuto

S

ST

T

PAC

iVAPS

ASV

ASVAuto

Range

Description

Settings

Circuit

Mask type

ü ü ü ü ü ü ü ü ü ü

Full Face / Nasal

Selects the type of mask used

 

 

/ Pillows / Nasal

by the patient.

 

 

Ultra / Pediatric

 

 

 

Default: Full Face

 

Tube type

ü ü ü ü ü ü ü ü ü ü

SlimLine /

Shows the type of air tubing

 

 

Standard / 3 m

used by the patient.

 

 

Default: Standard

Not displayed if ClimateLine or

 

 

 

ClimateLineMAX is connected

AB filter

ü ü ü ü ü ü ü ü ü ü

No / Yes

Enables or disables antibacterial

 

 

Default: No

filter.

 

 

Not displayed if H5i is

 

 

 

 

 

 

connected.

Ext.

ü ü ü ü ü ü ü ü ü ü

No / Yes

Enables or disables an external

humidifier

 

Default: No

humidifier.

 

 

 

Therapy

 

 

 

Mode

ü ü ü ü ü ü ü ü ü ü

CPAP / S / ST /

Sets the therapy mode available

 

 

T / PAC / iVAPS /

on the device.

 

 

ASVAuto / ASV /

 

 

 

AustoSet / VAuto

 

 

 

Default: CPAP

 

Set pressure

ü

4–20 cm H2O

Sets the fixed treatment

 

 

Default:

pressure.

 

 

 

 

 

4 cm H2O

 

EPR Level

ü ü

Off / 1 / 2 / 3

Sets the EPR value.

 

 

Default: 3

 

Height

ü

44-100 inches

Body height needed for the

 

 

(110–250 cm)

dead space determination.

 

 

Default: 70 inches

 

 

 

(175 cm)

 

Target Pt

ü

8–30 BPM

The rate input to the iVAPS

Rate

 

Default: 15 BPM

algorithm. This should be set at

 

 

the patient’s actual respiratory

 

 

 

 

 

 

rate.

Target Va

ü

1–30 L/min

Used to determine the amount

 

 

Default: 5.2 L/min

of pressure support required by

 

 

the iVAPS algorithm.

 

 

 

IPAP

ü ü ü ü

4–30 cm H2O

Sets the pressure which will be

 

 

Default:

delivered to the patient when

 

 

the device is triggered into

 

 

10 cm H2O

 

 

inspiration.

 

 

 

*Default settings may differ from country to country

 

 

16

Image 20
Contents S9 Vpap Tx Page S9 Vpap Tx Contents General warnings and cautions Welcome S9 Vpap Tx indications for useS9 Vpap Tx contraindications S9 Vpap Tx adverse effectsS9 Vpap Tx at a glance Masks Supplemental oxygenAntibacterial filters HumidifierTherapy mode Description Therapy modesS9 Vpap Tx basics Key FunctionSetting up the S9 Vpap Tx Filling the water tub English Air tubing Specifications Settings Mask and air tubing settingsMask type Settings Home menu Navigating the menusAbout the menus Changing settings via the Home menu Humidity levelClimate Control Mask FitViewing the treatment screens Treatment screen parameters Parameter Modes DescriptionSetup menu Setup menu accessSettings ConfigurationClinical setup menu parameters Parameter ModesRange Description Ü ü ü ü ü ü ü üParameter Modes Range Description Ü ü üLED Parameter Description ServiceInfo menu Temperature Managing Climate ControlMode Heated tubeDelivering therapy Adding supplemental oxygen To add supplemental oxygenCleaning and maintenance Filter EfficiencyTesting the alarms Working with alarmsAlarms menu Troubleshooting Problem/Possible cause SolutionNo display Insufficient air delivered from the deviceAlarm is activated and the LCD screen display disappears Water splashing on your face from the H5iLeaking water tub Air feels too warm/cold in the maskS9 Vpap Tx technical specifications Technical specificationsGeneral technical specifications Pneumatic flow path Value Range Display resolution Flow maximum at set pressuresDisplayed values MinSymbols ServicingLimited warranty ProductPage GLO Multi