ResMed S9 VPAP Tx manual Adding supplemental oxygen, To add supplemental oxygen

Page 26

Adding supplemental oxygen

WARNING

•• Explosion hazard—do not use in the vicinity of flammable anesthetics.

•• Oxygen supports combustion. Oxygen must not be used while smoking or in the presence of an open flame.

•• Always ensure that the device is turned on and airflow generated before the oxygen supply is turned on. Always turn the oxygen supply off before the device is turned off, so that unused oxygen does not accumulate within the device enclosure and create a risk of fire.

The S9 VPAP Tx is designed to be compatible with the following levels of supplemental oxygen:

•• up to 15 L/min for CPAP, S, T, ST, ASV and PAC modes

•• up to 4 L/min for AutoSet, VAuto, iVAPS and ASVAuto modes.

At a fixed rate of supplemental oxygen flow, the inhaled oxygen concentration will vary depending on:

•• where the oxygen is introduced;

•• the pressure settings;

•• patient breathing pattern;

•• mask selection; and

•• leak rate.

Adding oxygen may affect the trigger and cycle reliability, delivered pressure, and the accuracy of the displayed leak, tidal volume, minute ventilation and alveolar ventilation.

To add supplemental oxygen:

1.Fit the ResMed Oxygen Connector Port to the air outlet of the S9 VPAP Tx and fit the air tubing to the oxygen connector.

2.Fit the oxygen supply tubing to the port on the oxygen connector.

3.Attach the other end of the oxygen supply tubing to an oxygen flowmeter.

4.Optimize the therapy device settings first, and then add oxygen if baseline saturation remains low.

5.Titrate oxygen according to institutional guidelines.

6.Determine an initial oxygen flow rate during wakefulness. Only increase oxygen during sleep when titration of pressure is complete.

22

Image 26
Contents S9 Vpap Tx Page S9 Vpap Tx Contents General warnings and cautions S9 Vpap Tx contraindications WelcomeS9 Vpap Tx indications for use S9 Vpap Tx adverse effectsS9 Vpap Tx at a glance Antibacterial filters MasksSupplemental oxygen HumidifierTherapy mode Description Therapy modesS9 Vpap Tx basics Key FunctionSetting up the S9 Vpap Tx Filling the water tub English Air tubing Specifications Settings Mask and air tubing settingsMask type Settings Home menu Navigating the menusAbout the menus Climate Control Changing settings via the Home menuHumidity level Mask FitViewing the treatment screens Treatment screen parameters Parameter Modes DescriptionSettings Setup menuSetup menu access ConfigurationRange Description Clinical setup menu parametersParameter Modes Ü ü ü ü ü ü ü üParameter Modes Range Description Ü ü üLED Parameter Description ServiceInfo menu Mode TemperatureManaging Climate Control Heated tubeDelivering therapy Adding supplemental oxygen To add supplemental oxygenCleaning and maintenance Filter EfficiencyTesting the alarms Working with alarmsAlarms menu No display TroubleshootingProblem/Possible cause Solution Insufficient air delivered from the deviceLeaking water tub Alarm is activated and the LCD screen display disappearsWater splashing on your face from the H5i Air feels too warm/cold in the maskS9 Vpap Tx technical specifications Technical specificationsGeneral technical specifications Pneumatic flow path Displayed values Value Range Display resolutionFlow maximum at set pressures MinSymbols ServicingLimited warranty ProductPage GLO Multi