Problem/Possible cause | Solution |
Air filter is blocked. | Replace your air filter. Disconnect the power cord and then |
| reconnect it to restart the device. |
Air tubing is blocked. | Check your air tubing and remove any blockages. Disconnect |
| the power cord and then reconnect it to restart the device. |
Humidity level setting is too high, | Turn the humidity level setting down and empty the water from |
resulting in accumulation of water in | the air tubing. |
the air tubing. |
|
Displays message: Check ResMed 30/90W Power Supply Unit and fully insert the connector | |
The DC plug is partially inserted into | Fully insert the DC plug. |
the back of the device or inserted |
|
too slowly. |
|
A | Remove the power supply unit and replace with a ResMed |
connected to the device. | power supply unit. |
The power supply unit is being | Make sure that the power supply unit is free from bedding, |
covered by bedding. | clothes or other objects that could cover it. |
Water splashing on your face from the H5i | |
The water tub is overfilled. | Check that the water level is below the maximum water level |
| mark. |
Condensation is forming in the air | Turn the humidity level setting down via the menu. |
tubing and mask. |
|
Leaking water tub |
|
The water tub may be damaged or | Contact your service provider for a replacement. |
cracked. |
|
The cleanable water tub is not | Check for damage and reassemble the cleanable water tub |
assembled correctly. | correctly. |
Air feels too warm/cold in the mask
The temperature of the ClimateLine or ClimateLineMAX heated air tubing is set too high/low.
Turn up/down the heated air tubing temperature via the menu.
Alarm is activated and the LCD screen display disappears
Power failure. | Remove the patient’s mask until power is restored. |
Power cord is disconnected or mains | Ensure the power cord is connected and the mains power |
power switch is turned off during | outlet (if available) is on. |
therapy. |
|
Displays message: Tube blocked, please check your tube
Air tubing is blocked.
Check the air tubing and remove any blockages. Disconnect the power cord and then reconnect it to restart the device.
Displays message: Alarm module fault, please contact service provider
General failure of the device and/or the alarm module. Therapy cannot be started again.
Contact your service provider immediately.
26