ResMed S9 VPAP Tx manual Water splashing on your face from the H5i, Leaking water tub

Page 30

Problem/Possible cause

Solution

Air filter is blocked.

Replace your air filter. Disconnect the power cord and then

 

reconnect it to restart the device.

Air tubing is blocked.

Check your air tubing and remove any blockages. Disconnect

 

the power cord and then reconnect it to restart the device.

Humidity level setting is too high,

Turn the humidity level setting down and empty the water from

resulting in accumulation of water in

the air tubing.

the air tubing.

 

Displays message: Check ResMed 30/90W Power Supply Unit and fully insert the connector

The DC plug is partially inserted into

Fully insert the DC plug.

the back of the device or inserted

 

too slowly.

 

A non-ResMed power supply unit is

Remove the power supply unit and replace with a ResMed

connected to the device.

power supply unit.

The power supply unit is being

Make sure that the power supply unit is free from bedding,

covered by bedding.

clothes or other objects that could cover it.

Water splashing on your face from the H5i

The water tub is overfilled.

Check that the water level is below the maximum water level

 

mark.

Condensation is forming in the air

Turn the humidity level setting down via the menu.

tubing and mask.

 

Leaking water tub

 

The water tub may be damaged or

Contact your service provider for a replacement.

cracked.

 

The cleanable water tub is not

Check for damage and reassemble the cleanable water tub

assembled correctly.

correctly.

Air feels too warm/cold in the mask

The temperature of the ClimateLine or ClimateLineMAX heated air tubing is set too high/low.

Turn up/down the heated air tubing temperature via the menu.

Alarm is activated and the LCD screen display disappears

Power failure.

Remove the patient’s mask until power is restored.

Power cord is disconnected or mains

Ensure the power cord is connected and the mains power

power switch is turned off during

outlet (if available) is on.

therapy.

 

Displays message: Tube blocked, please check your tube

Air tubing is blocked.

Check the air tubing and remove any blockages. Disconnect the power cord and then reconnect it to restart the device.

Displays message: Alarm module fault, please contact service provider

General failure of the device and/or the alarm module. Therapy cannot be started again.

Contact your service provider immediately.

26

Image 30
Contents S9 Vpap Tx Page S9 Vpap Tx Contents General warnings and cautions S9 Vpap Tx contraindications WelcomeS9 Vpap Tx indications for use S9 Vpap Tx adverse effectsS9 Vpap Tx at a glance Antibacterial filters MasksSupplemental oxygen HumidifierTherapy mode Description Therapy modesS9 Vpap Tx basics Key FunctionSetting up the S9 Vpap Tx Filling the water tub English Mask and air tubing settings Mask type SettingsAir tubing Specifications Settings Navigating the menus About the menusHome menu Climate Control Changing settings via the Home menuHumidity level Mask FitViewing the treatment screens Treatment screen parameters Parameter Modes DescriptionSettings Setup menuSetup menu access ConfigurationRange Description Clinical setup menu parametersParameter Modes Ü ü ü ü ü ü ü üParameter Modes Range Description Ü ü üLED Service Info menuParameter Description Mode TemperatureManaging Climate Control Heated tubeDelivering therapy Adding supplemental oxygen To add supplemental oxygenCleaning and maintenance Filter EfficiencyWorking with alarms Alarms menuTesting the alarms No display TroubleshootingProblem/Possible cause Solution Insufficient air delivered from the deviceLeaking water tub Alarm is activated and the LCD screen display disappearsWater splashing on your face from the H5i Air feels too warm/cold in the maskTechnical specifications General technical specificationsS9 Vpap Tx technical specifications Pneumatic flow path Displayed values Value Range Display resolutionFlow maximum at set pressures MinSymbols ServicingLimited warranty ProductPage GLO Multi