ResMed S9 VPAP Tx manual Technical specifications, General technical specifications

Page 31

Technical specifications

General technical specifications

Power supply

90W power supply unit

 

Input range: 100–240V; 50–60Hz; 115V, 400Hz nominal for aircraft use

 

Typical power consumption: 70W (80VA)

 

Maximum power consumption: 110W (120VA)

 

90W DC/DC converter

 

Nominal inputs: 12V, 24V

 

Typical power consumption: 70W

 

Maximum power consumption: 110W

Environmental

Operating temperature: +41ºF to +95ºF (+5ºC to +35ºC)

conditions

Note: The air flow for breathing produced by this therapy device can be higher

 

than the temperature of the room. Under extreme ambient temperature conditions

 

(104°F/40°C) the device remains safe.

 

Operating humidity:10 to 95% non-condensing

 

Operating altitude: Sea level to 8,500’ (2,591 m); air pressure range 1013 hPa to

 

738 hPa

 

Storage and transport temperature: -4ºF to +140ºF (-20ºC to +60ºC)

 

Storage and transport humidity: 10 to 95% non-condensing

Electromagnetic

Product complies with all applicable electromagnetic compatibility requirements

compatibility

(EMC) according to IEC60601-1-2, for residential, commercial, and light industry

 

environments.

 

It is recommended that mobile communication devices are kept at least one

 

meter away from the device.

 

Information regarding the electromagnetic emissions and immunity of these

 

ResMed devices can be found on www.resmed.com, on the Products page under

 

Service and Support. Click on the PDF file for your language.

Aircraft use

ResMed confirms that the S9 VPAP Tx meets the Federal Aviation Administration

 

(FAA) requirements (RTCA/DO-160, section 21, category M) for all phases of air

 

travel.

IEC 60601-1

Class II (double insulation), Type BF, Ingress Protection IP21

classification

 

Measuring and

Pressure sensor: Internally located at device outlet, analog gauge pressure type,

display devices

-5 to +45 cm H2O

 

Flow sensor: Internally located at device inlet, digital mass flow type,

 

-70 to +200 L/min

 

S9 VPAP Tx technical specifications

Mode pressure

CPAP, AutoSet mode

ranges

Pressure: 4–20 cm H2O;

 

S, ST, T and PAC mode

 

IPAP: 4–30 cm H2O; EPAP: 2–25 cm H2O;

 

iVAPS mode

 

PS: 0–27 cm H2O; EPAP: 2–25 cm H2O;

 

VAuto mode

 

PS: 0-10 cm H2O; Min EPAP: 4 cm H2O; Max IPAP: 25 cm H2O;

 

ASV and ASVAuto mode

 

PS: 0–20 cm H2O; EPAP: 4–15 cm H2O;

Maximum single

Maximum single fault steady state pressure: 30 cm H2O—if pressure exceeded

fault pressure

for > 6 sec; 40 cm H2O—if pressure exceeded for >1 sec

27 English

Image 31
Contents S9 Vpap Tx Page S9 Vpap Tx Contents General warnings and cautions S9 Vpap Tx adverse effects WelcomeS9 Vpap Tx indications for use S9 Vpap Tx contraindicationsS9 Vpap Tx at a glance Humidifier MasksSupplemental oxygen Antibacterial filtersTherapy modes Therapy mode DescriptionKey Function S9 Vpap Tx basicsSetting up the S9 Vpap Tx Filling the water tub English Mask type Settings Mask and air tubing settingsAir tubing Specifications Settings About the menus Navigating the menusHome menu Mask Fit Changing settings via the Home menuHumidity level Climate ControlViewing the treatment screens Parameter Modes Description Treatment screen parametersConfiguration Setup menuSetup menu access SettingsÜ ü ü ü ü ü ü ü Clinical setup menu parametersParameter Modes Range DescriptionÜ ü ü Parameter Modes Range DescriptionLED Info menu ServiceParameter Description Heated tube TemperatureManaging Climate Control ModeDelivering therapy To add supplemental oxygen Adding supplemental oxygenFilter Efficiency Cleaning and maintenanceAlarms menu Working with alarmsTesting the alarms Insufficient air delivered from the device TroubleshootingProblem/Possible cause Solution No displayAir feels too warm/cold in the mask Alarm is activated and the LCD screen display disappearsWater splashing on your face from the H5i Leaking water tubGeneral technical specifications Technical specificationsS9 Vpap Tx technical specifications Pneumatic flow path Min Value Range Display resolutionFlow maximum at set pressures Displayed valuesServicing SymbolsProduct Limited warrantyPage GLO Multi