Mask and air tubing settings
Use the following settings below for each mask type:
Mask type | Settings |
Full Face | Full Face |
Pillows | Pillows |
Nasal | Nasal (for Ultra Mirage mask, use ‘Nasal Ultra’) |
Pediatric | Pediatric |
Notes:
•• For more information on assembling the mask see the mask user guide.
•• For a complete list of recommended masks and their settings go to www.resmed.com on the Products page under Service & Support. If you do not have internet access, please contact your ResMed representative.
The S9 VPAP Tx is compatible with the following air tubing:
Air tubing | Specifications | Settings |
ClimateLine | Heated | Automatically detected |
| Length: 6’6” (2 m) |
|
| Inner diameter: 0.6” (15 mm) |
|
ClimateLineMAX | Heated | Automatically detected |
| Length: 6’3” (1.9 m) |
|
| Inner diameter: 0.75” (19 mm) |
|
SlimLine | Length: 6’ (1.8 m) | If using the SlimLine, Standard or 3 m |
| Inner diameter: 0.6” (15 mm) | air tubing, adjust the tube setting via the |
Standard | Length: 6’6” (2 m) | Setup menu. |
| ||
| Inner diameter: 0.75” (19 mm) |
|
3 m | Length: 9’10” (3 m) |
|
| ||
| Inner diameter: 0.75” (19 mm) |
|
Notes:
•• When using the SlimLine or ClimateLine above 20 cm H2O, the device optimum performance may not be reached if used with an antibacterial filter. The device performance must be checked prior to using the SlimLine for use with an antibacterial filter.
•• The ClimateLine/ClimateLineMAX are designed only for use with the H5i.
10