ResMed S9 VPAP Tx manual Parameter Modes Range Description, Ü ü ü

Page 21

Parameter

 

Modes

 

 

 

 

Range

Description

CPAP

AutoSet

VAuto S ST T

PAC

iVAPS

ASV

ASVAuto

 

 

EPAP

 

ü ü ü ü ü ü

 

3–25 cm H2O**

Sets the pressure which will

 

 

 

 

 

 

 

Default:

be delivered to the patient

 

 

 

 

 

 

 

when the device is cycled into

 

 

 

 

 

 

 

4 cm H2O

 

 

 

 

 

 

 

expiration.

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

Dependent on IPAP.

Min PS

 

 

 

ü ü ü

0–20 cm H2O

Minimum pressure support in

 

 

 

 

 

 

 

Default:

iVAPS mode.

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

4 cm H2O

Dependent on EPAP.

Max PS

 

 

 

ü ü ü

0–28 cm H2O

Maximum pressure support in

 

 

 

 

 

 

 

Default:

iVAPS mode.

 

 

 

 

 

 

 

20 cm H2O

Dependent on EPAP and Min

 

 

 

 

 

 

 

 

PS.

Respiratory

 

ü ü ü

 

 

 

5–50 BPM

Sets the breaths per minute

Rate

 

 

 

 

 

 

Default: 10 BPM

(BPM) or ‘backup’ rate.

 

 

 

 

 

 

 

 

Ti Max

 

ü ü ü

 

ü

 

 

0.3–4.0 sec

Sets the maximum limit on the

 

 

 

 

 

 

 

Default: 2 sec

time the device spends in IPAP.

 

 

 

 

 

 

 

Dependent on Respiratory Rate.

 

 

 

 

 

 

 

 

Ti Min

 

ü ü ü

 

ü

 

 

0.1–4.0 sec

Sets the minimum limit on the

 

 

 

 

 

 

 

Default: 0.3 sec

time the device spends in IPAP.

 

 

 

 

 

 

 

Dependent on Ti Max.

 

 

 

 

 

 

 

 

Ti

 

ü ü

 

 

 

0.1–4.0 sec

Sets the duration of inspiration

 

 

 

 

 

 

 

Default: 2 sec

in timed breath.

 

 

 

 

 

 

 

Dependent on Respiratory Rate.

 

 

 

 

 

 

 

 

Rise Time

 

ü ü ü ü ü

 

 

Min /

Sets the time taken for the

 

 

 

 

 

 

 

100–900 ms

device to reach to IPAP.

 

 

 

 

 

 

 

Default: 150 ms

Dependent on Ti Max and Ti.

Trigger

 

ü ü ü

ü ü

 

 

Very Low / Low /

Sets the level of inspiratory flow

 

 

 

 

 

 

 

Med / High / Very

above which the device changes

 

 

 

 

 

 

 

High

from EPAP to IPAP.

 

 

 

 

 

 

 

Default: Med

 

Cycle

 

ü ü ü

 

ü

 

 

Very Low / Low /

Sets the level of inspiratory flow

 

 

 

 

 

 

 

Med / High / Very

below which the device changes

 

 

 

 

 

 

 

High

from IPAP to EPAP.

 

 

 

 

 

 

 

Default: Med

 

Max IPAP

 

ü

 

 

 

 

4–25 cm H2O

Sets the maximum inspiratory

 

 

 

 

 

 

 

Default:

pressure delivered by the

 

 

 

 

 

 

 

device.

 

 

 

 

 

 

 

25 cm H2O

 

 

 

 

 

 

 

 

Min EPAP

 

ü

 

 

 

 

4–[Max IPAP]

Sets the minimum expiratory

 

 

 

 

 

 

 

cm H2O

pressure delivered by the

 

 

 

 

 

 

 

Default:

device.

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

4 cm H2O

 

Pressure

 

ü

 

 

 

 

0–10 cm H2O

Difference between IPAP and

Support

 

 

 

 

 

 

Default:

EPAP. Adjust for patient comfort.

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

4 cm H2O

 

** 2-25 cm H2O for Europe and APAC

 

 

 

 

 

 

17 English

Image 21
Contents S9 Vpap Tx Page S9 Vpap Tx Contents General warnings and cautions S9 Vpap Tx indications for use WelcomeS9 Vpap Tx contraindications S9 Vpap Tx adverse effectsS9 Vpap Tx at a glance Supplemental oxygen MasksAntibacterial filters HumidifierTherapy modes Therapy mode DescriptionKey Function S9 Vpap Tx basicsSetting up the S9 Vpap Tx Filling the water tub English Mask and air tubing settings Mask type SettingsAir tubing Specifications Settings Navigating the menus About the menusHome menu Humidity level Changing settings via the Home menuClimate Control Mask FitViewing the treatment screens Parameter Modes Description Treatment screen parametersSetup menu access Setup menuSettings ConfigurationParameter Modes Clinical setup menu parametersRange Description Ü ü ü ü ü ü ü üÜ ü ü Parameter Modes Range DescriptionLED Service Info menuParameter Description Managing Climate Control TemperatureMode Heated tubeDelivering therapy To add supplemental oxygen Adding supplemental oxygenFilter Efficiency Cleaning and maintenanceWorking with alarms Alarms menuTesting the alarms Problem/Possible cause Solution TroubleshootingNo display Insufficient air delivered from the deviceWater splashing on your face from the H5i Alarm is activated and the LCD screen display disappearsLeaking water tub Air feels too warm/cold in the maskTechnical specifications General technical specificationsS9 Vpap Tx technical specifications Pneumatic flow path Flow maximum at set pressures Value Range Display resolutionDisplayed values MinServicing SymbolsProduct Limited warrantyPage GLO Multi