ResMed User Manual for CPAP and VPAP Devices: Alarms and Testing Guide

Page 28

Working with alarms

The S9 VPAP Tx is fitted with an alarm module that continuously monitors both therapy and device conditions. Alarms are only activated when therapy is running. An alarm condition can be indicated by an audible sound, a flashing yellow LED or a message on the screen.

Note: The alarms should be tested after final system setup to ensure the alarms work as required.

Alarms menu

The alarms pre-set for the device are:

•• Power fail

•• Blocked tube*

•• System fault (system error).

* Blocked tube alarm is only triggered reliably for pressures above 10 cm H2O.

If the system has not been properly assembled, the device will trigger an alarm. Check that the air tubing has been properly attached to the device, mask and humidifier.

Clearing the alarm message

When an alarm activates, a corresponding alarm message is displayed. To clear an alarm message, press . This allows you to return to the previously displayed screen.

Note: Alarms are cleared by pressing .

Testing the alarms

When the device is turned on, the alarm LED will flash and the alarm will sound to confirm that the alarm is working.

To test the alarm manually or to adjust the alarm volume using the menus, go to the Settings menu and select Alarm Vol/Test. When the setting is selected and confirmed, the alarm will sound and all LEDs will be active.

The alarms should be tested weekly. To test some of the alarm conditions, follow the procedures described below. When completed, stop therapy and return all settings to their original settings appropriate to the patient before delivering therapy.

Power fail

1.Set up the device with the air tubing attached, but no mask.

2.Press the Start/Stop key to start therapy.

3.Unplug the DC plug of the power supply unit from the rear of the device. The alarm activates immediately.

4.Plug the DC plug back in. The alarm stops.

Blocked tube

1.Set up the device with the air tubing attached, but no mask.

2.Set pressure above 12 cm H2O in CPAP mode.

Note: Blocked Tube alarm only activates above 10 cm H2O.

3.Press the Start/Stop key to start therapy.

4.Block the air tubing with your hand. The alarm activates when tubing is blocked for 30–50 seconds.

5.Unblock the air tubing. The alarm stops.

System Fault

No testing for system fault.

24

Image 28
Contents S9 Vpap Tx Page S9 Vpap Tx Contents General warnings and cautions Welcome S9 Vpap Tx indications for useS9 Vpap Tx contraindications S9 Vpap Tx adverse effectsS9 Vpap Tx at a glance Masks Supplemental oxygenAntibacterial filters HumidifierTherapy mode Description Therapy modesS9 Vpap Tx basics Key FunctionSetting up the S9 Vpap Tx Filling the water tub English Mask type Settings Mask and air tubing settingsAir tubing Specifications Settings About the menus Navigating the menusHome menu Changing settings via the Home menu Humidity levelClimate Control Mask FitViewing the treatment screens Treatment screen parameters Parameter Modes DescriptionSetup menu Setup menu accessSettings ConfigurationClinical setup menu parameters Parameter ModesRange Description Ü ü ü ü ü ü ü üParameter Modes Range Description Ü ü üLED Info menu ServiceParameter Description Temperature Managing Climate ControlMode Heated tube Delivering therapy Adding supplemental oxygen To add supplemental oxygenCleaning and maintenance Filter EfficiencyAlarms menu Working with alarmsTesting the alarms Troubleshooting Problem/Possible cause SolutionNo display Insufficient air delivered from the deviceAlarm is activated and the LCD screen display disappears Water splashing on your face from the H5iLeaking water tub Air feels too warm/cold in the maskGeneral technical specifications Technical specificationsS9 Vpap Tx technical specifications Pneumatic flow path Value Range Display resolution Flow maximum at set pressuresDisplayed values MinSymbols ServicingLimited warranty ProductPage GLO Multi