Graco 249881, 246008 instruction manual Replacing the Meter Electronics

Page 28

Service

Replacing the Meter Electronics

Electronics Repair Kits 246675 for meter 246008 and 249882 for meter 249881 are available for replacing the Matrix meter electronics. See the Parts List for the 246008 and 249881 Electronic Meters on page 33. The parts with a † next to their reference numbers are included in the Electronics Repair Kit.

When repairing the meter using the Electronics Repair kit be sure to use all of the new parts. To replace the electronics do the following.

The PSM switch (101) comes attached to the right meter cover half (18). Do not remove it during the meter electronics repair procedure.

CAUTION

Be sure to use the disposable grounding wrist strap included with the repair kit. Static electricity can dam- age the electrical components within the meter dur- ing the repair.

1.Relieve the Pressure.

WARNING

To reduce the risk of serious injury whenever you are instructed to relieve pressure, always follow the Pressure Relief Procedure on page 8.

1.Remove the battery (11). See Parts Drawing for the 246008 and 249881 Electronic Meters on page 32.

2.Slide the rear boot (24) off the hose.

3.Remove the meter from the fluid hose.

4.Remove the impact guard (40) from the meter.

5.With a phillips screwdriver remove the four phillips screws (42) holding the bezel to the cover halves (18 & 19). See FIG. 41.

6.With a T-10 torx driver remove the remaining eleven screws (12) holding the cover halves together (18 & 19).

12

42 (both sides)

FIG. 41

7.Pull apart the two cover halves leaving all parts in the right cover half (18).

All services to the meter should be done using the right cover half (18) of the meter. If parts should disen- gage as the covers are separated, reposition the parts back into the right cover half (18). See FIG. 42.

19

18

FIG. 42

8.Place the new right cover (18) along side of the meter.

9.Remove the trigger assembly (13, 14, 22, 23, 36, & 37) from the old cover (18) and set aside.

10.Remove the meter housing from the old right cover (18).

11.Disconnect the cables/wires (A, B, C, & D) that con- nect the bezel (31) to the circuit board (112) and the valve housing to the circuit board. See FIG. 43.

28

309497L

Image 28
Contents Matrix Meter Contents Manual ConventionsSkin Injection Hazard Programming the Meter Meter ProgrammingProgramming the Meter Meter Installation DESCRIPTION, See F IGMounting Brackets DESCRIPTION, See FIGPre-installation Procedure Pressure Relief ProcedureGrounding Meter Installation to Hose Installing Nozzle ExtensionPressurizing the System Asleep / Awake Mode High/Low Flow Rate IndicatorBattery Life Indicator Meter OperationParts Room Authorization Security ModesSystem Monitoring PIN CodeWork Orders and Job Numbers Assigning Work Orders and Job Numbers from the PCAssigning Work Orders and Job Numbers at the Meter 32567 IC9 2AC03 458A1 2233APlacing a Meter in the Emergency Mode Emergency Mode1222 Manual Dispense Mode Preset Dispense ModeDispense Options Restricted Preset Dispense Mode Removing and Replacing the BatteryQuick-LockNozzle Nozzle opens and closes with Less than one turnTroubleshooting Problem Cause SolutionFications Flow. Warning do not exceed the maximum workingPressure of the meter 1500 psi 103 bar. If flushing Problem Cause Solution Metric Measurement Calculating the Calibration FactorMeasurement 309497L Service Replacing or Cleaning FilterFollow the Pressure Relief Procedure, page 8, whenever you Replacing Quick-lockNozzleReplacing the Bezel Service Replacing the Trigger Relieve the Pressure Ti4323 Service Replacing the Meter Electronics Service Service Service Parts Drawing for the 246008 and 249881 Electronic Meters Parts List for the 246008 and 249881 Electronic Meters Description QtyNozzle Extension Accessories KitQty DescriptionTechnical Specifications 309497L Dimensions Graco Standard Warranty Graco Phone Numbers