A&D Ambulatory Blood Pressure Monitor Automatic measurement, Stopping a measurement, Memory

Page 11

Automatic measurement

This measurement starts or stops using the AUTO ON/OFF key. When this measurement is started, the recorder begins to work in accordance with preset time intervals from the preset time of the internal clock. Refer to "Selection for the automatic measurement".

In the automatic measurement, an "A" appears in the upper left of the display.

The recorder automatically measures the patient's blood pressure at the time that is pointed out by "the frequency" and "the start time" (by the programed time intervals).

When a measurement error occurs and there is 10 minutes until the next measurement, the measurement is retried after approx. 30 seconds.

If a measurement is retried, only the data from the retry is saved.

The recorder automatically adjusts the proper pressure, exhaust velocity and end of measurement.

Refer to "Selection for the automatic measurement" and "Automatic measure- ment by programed time intervals" about operation and entering parameters.

Stopping a measurement

If you press the

START STOP

key during a measurement, the recorder exhausts

the air and stops the measurement.

Concealing the measurement value

This function works only while using automatic measurement.

This function does not display the SYS, DIA or pulse rate for the automatic measurement, but the data is saved in memory.

This function can select "reveal" or "conceal" at "Parameters for the display and clock". Refer to this section.

If you select "conceal", the recorder displays the clock during a measurement.

If you reset the recorder, this parameter is set to "reveal".

Pressurization

The pressure is automatically selected by the recorder while in the automatic measurement mode.

The first pressure is set to approx. 185 mmHg. This value automatically adjusts to the proper value after the first measurement. If the first inflation is not successful, the recorder retries twice.

If you reset the recorder, the first inflation value is reset to 185 mmHg.

Memory

The TM-2430 recorder can store 300 sets of data.

A data set consists of a SYS, a DIA and a pulse rate.

When memory becomes full, the recorder displays an "M". Until you clear the data, you can not measure blood pressure.

When the recorder saves data for more than one patient, data management becomes complication. We recommend that patient data is transferred and cleared from memory.

When a "B" is displayed, the backup batteries that is preserving a patient`s data are weak. Please replace batteries as soon as possible.

ID number

If you reset the recorder, the ID number is set to "1".

The ID number can be set using the optional software.

Specifications

Page 9

Functions and Specifications

Image 11
Contents TM-2430 Page Contents Definitions Before useCompliance Precautions for Use During use Periodic maintenanceStorage Before usePacking List and Component Names WelcomeProduct overview Welcome and Introduction Air socket Display SymbolsFunctions and Specifications SpecificationsFeatures Pressurization Automatic measurementStopping a measurement Concealing the measurement valuePerformance specifications Environment specifications Steps for replacing the batteries Activating the recorderReplacing Batteries Display & key Conditions When Recorder Switched OnWhen recorder is Condition Actions Turned on Parameters for the Display and ClockSteps for setting the display and clock Mode Steps for selecting a modeSelection for the Automatic Measurement Set up procedure Mode III SettingsAs the start time of the third block Steps for automatic measurementFrom previous Steps for deleting old data Resetting the RecorderSteps for reset Deleting Old DataManual measurement Stopping or canceling an automatic measurementPreparing the patient Patient instructionsSteps for attaching the cuff and recorder Use of the cuff coverAttaching the cuff and recorder Range, you need a proper cuff Preparation of the Carrying Case Using the BeltUsing the Shoulder Band To stop a Current Measurement OperationAutomatic Measurement Manual MeasurementData Transmission to a Printer Specifications for an adaptable printerData Transfer Print sample Steps for data transmissionData Transmission to a Computer Using Analysis Soft Analysis Software and Communication Cables Options and accessoriesOthers Cuffs and Other AccessoriesCuff sleeves Cuff coverSteps for checking accuracy MaintenanceChecking Accuracy Required equipmentPeriodical inspection Cleaning the cuff and recorderProblem solving Error code updates without announcement Error codesMaintenance Error Code Memo