A&D Understanding Blood Pressure Monitor Error Codes for Devices

Page 34

Error codes

Caution

The error code updates without announcement.

Error

Meaning

Status

Operation and Treatment

code

 

 

 

 

 

E00

No clock

All parameters are lost.

Enter clock parameters. Refer to

 

parameter

Reset status.

"Setup of display and clock"

E03

Pressure zero

An error code is displayed

Release the air from the cuff

 

error

 

without cuff inflation.

completely.

E04

Low battery

Measurement is stopped.

Replace with new batteries.

 

 

 

An error code is displayed.

Restart the auto mode if you use it.

 

 

 

Auto mode is quit.

 

E05

Inflation error

Inflation pressure does not

Wrap the cuff and connect to main

 

 

 

reach the target pressure.

unit exactly. If you can not clear the

 

 

 

 

 

error, there may be an air leak and

 

 

 

 

 

repair is necessary

E06

Above 320mmHg

An error code is displayed.

Do not move and try to relax during

 

 

 

 

 

the measurement. If you can not

 

 

 

 

 

clear the error, the product will

 

 

 

 

 

requier repair.

E07

Controlled stop

Air is exhausted. An error

Do not press the STOP key if you do

 

using STOP key

code is displayed.

not need to use it.

E08

Pulsation can not

Measurable pulsation is

Do not move and try to relax during

 

be measured

searched to 20mmHg in

the measurement. The error occurs

 

 

 

constant exhaust. An error

when measurable pulsations are not

 

 

 

code is displayed.

received due to thick cloth or quick

 

 

 

 

 

motion.

E10

Pulsations can not In the measurement, Quick

Do not move and try to relax during

 

be detected

exhaust is executed. An

the measurement.

 

because the

error code is displayed.

 

 

patient may have

 

 

 

 

moved.

 

 

 

 

E20

Pulse rate < 30

An error code is displayed.

Measure the blood pressure by other

 

200 < Pulse rate

 

 

methods.

E21

DIA < 40

 

 

 

 

160 < DIA

DIA

: Diastolic Blood Pressure

E22

SYS < 60

SYS

: Systolic Blood Pressure

 

280 < SYS

 

DSD : The Difference between Systolic Blood Pressure

E23

DSD < 10

 

: and Diastolic Blood Pressure.

 

150 < DSD

 

 

 

 

 

Error Code

 

 

Page 32

Maintenance

Image 34
Contents TM-2430 Page Contents Compliance Before useDefinitions Precautions for Use Before use Periodic maintenanceStorage During useWelcome and Introduction WelcomeProduct overview Packing List and Component NamesAir socket Symbols DisplayFeatures SpecificationsFunctions and Specifications Concealing the measurement value Automatic measurementStopping a measurement PressurizationPerformance specifications Environment specifications Replacing Batteries Activating the recorderSteps for replacing the batteries Parameters for the Display and Clock Conditions When Recorder Switched OnWhen recorder is Condition Actions Turned on Display & keySteps for setting the display and clock Selection for the Automatic Measurement Steps for selecting a modeMode Mode III Settings Set up procedureSteps for automatic measurement As the start time of the third blockFrom previous Deleting Old Data Resetting the RecorderSteps for reset Steps for deleting old dataPatient instructions Stopping or canceling an automatic measurementPreparing the patient Manual measurementAttaching the cuff and recorder Use of the cuff coverSteps for attaching the cuff and recorder Range, you need a proper cuff Using the Shoulder Band Using the BeltPreparation of the Carrying Case Manual Measurement OperationAutomatic Measurement To stop a Current MeasurementData Transfer Specifications for an adaptable printerData Transmission to a Printer Steps for data transmission Print sampleData Transmission to a Computer Using Analysis Soft Options and accessories Analysis Software and Communication CablesCuff cover Cuffs and Other AccessoriesCuff sleeves OthersRequired equipment MaintenanceChecking Accuracy Steps for checking accuracyProblem solving Cleaning the cuff and recorderPeriodical inspection Error codes Error code updates without announcementMaintenance Error Code Memo

Ambulatory Blood Pressure Monitor specifications

The A&D Ambulatory Blood Pressure Monitor is a cutting-edge device designed for the accurate measurement of blood pressure over an extended period. This innovative tool is particularly beneficial for diagnosing hypertension and evaluating blood pressure variability, making it essential for both healthcare professionals and patients.

One of the main features of the A&D Ambulatory Blood Pressure Monitor is its ability to perform measurements during the patient’s normal daily activities. This is crucial for capturing accurate readings, as blood pressure can fluctuate significantly throughout the day due to various factors such as stress, physical activity, and diet. By documenting blood pressure readings at set intervals over 24 hours, this monitor provides a comprehensive overview of the patient's cardiovascular health.

The device utilizes advanced oscillometric technology, which enhances accuracy by analyzing the oscillations in the arterial wall during cuff inflation. This method is recognized for its reliability compared to traditional auscultatory techniques. The monitor is equipped with a large memory capacity, allowing it to store numerous readings which can be analyzed later. This feature is invaluable for doctors who need to review the patient’s blood pressure patterns over time.

User-friendliness is another hallmark of the A&D Ambulatory Blood Pressure Monitor. The device typically features an intuitive display and controls that make it easy to navigate through settings and review stored data. Many models offer wireless connectivity, enabling seamless data transfer to smartphones or computers for further analysis and reporting. This capability aids healthcare providers in timely decision-making regarding treatment options.

Additionally, the A&D monitor is designed with comfort in mind. It comes with a soft, adjustable cuff that can accommodate a range of arm sizes, ensuring an accurate fit and comfortable experience for the patient. The device is ergonomically designed and lightweight, making it easy to wear throughout the day without interfering with daily activities.

The A&D Ambulatory Blood Pressure Monitor not only promotes better patient compliance but also enhances clinical outcomes by enabling continuous and accurate monitoring of blood pressure in real-world settings. Overall, this device embodies the convergence of technology and healthcare, providing invaluable data that fosters informed decisions and improves patient care.