Error | Meaning | Status |
| Operation and Treatment |
| ||||
code |
| |||
|
|
|
| |
E30 | Measurement is | Air is exhausted from the |
| Repair is necessary because of slow |
| greater than 120 | cuff, and an error code is |
| inflation or slow constant exhaust. |
| seconds. | displayed. |
|
|
|
|
|
|
|
E31 | The constant | Air is exhausted from the |
| Repair is necessary because of slow |
| exhaust is greater | cuff, and an error code is |
| constant exhaust. |
| than 60 seconds. | displayed. |
|
|
|
|
|
|
|
E32 | Clock error. | An error code is displayed. |
| If you not clear this error, the product |
|
|
|
| needs repair. |
|
|
|
|
|
E50 | Pressure offset | An error code is displayed |
| Release the air from the cuff com- |
| error to measure | at restarting the product. |
| pletely, reset the product. If you not |
| pulsation. |
|
| clear this error, repair is necessary. |
|
|
|
|
|
E52 | Memory error. | An error code is displayed |
| The product needs repair. |
|
| at restarting the product. |
|
|
|
|
|
|
|
E53 | Battery contact is | The measurement is |
| Replace batteries correctly. If you can |
| defective. | stopped, air is released |
| not clear this error, the product needs |
|
| from the cuff and an error |
| repair. |
|
| code is displayed. |
|
|
|
|
|
|
|
E55 | Exhaust error. | An error code is displayed |
| Relax and do not move during the |
E56 |
| at measurement. |
| measurement. If this error occurs |
E57 |
|
|
| many times, repair is necessary. |
|
|
|
|
|
E60 | Interval setting | Start time is not proper, |
| Enter parameters for the interval |
| error. | interval of last block is not |
| correctly. |
|
| set in the unit of 120 min. |
|
|
|
|
|
|
|
E70 | The error code is displayed |
| ||
E71 |
| during communications. |
| If you can not clear this error, the |
E72 |
|
|
| product needs repair. |
E73 |
|
|
|
|
|
|
|
|
|
E74 | Low battery for |
|
| Replace batteries with new ones and |
| communication. |
|
| restart communication. |
|
|
|
|
|
E75 | Protcol error due |
|
| |
| to external |
|
| If you can not clear this error, the |
| equipment. |
|
| product needs repair. |
|
|
|
|
|
E90 | Pressure zero | This error code is |
| Release the air from the cuff |
| error for safety | displayed before the |
| completely. |
| circuit. | measurement. |
|
|
|
|
|
|
|
E91 | Safety circuit | Patient moved during the |
| Relax and try to quiet during the |
| detects over load | measurement. |
| measurement. If it occurs in quiet, the |
| pressure. |
|
| recorder needs repair. |
|
|
|
|
|
Other |
| Monitor code is displayed. |
| Reset. Turn on power switch again. |
|
|
|
|
|
Maintenance | Page 33 | Error Code |