A&D Ambulatory Blood Pressure Monitor instruction manual Maintenance Error Code

Page 35

Error

Meaning

Status

 

Operation and Treatment

 

code

 

 

 

 

 

E30

Measurement is

Air is exhausted from the

 

Repair is necessary because of slow

 

greater than 120

cuff, and an error code is

 

inflation or slow constant exhaust.

 

seconds.

displayed.

 

 

 

 

 

 

 

E31

The constant

Air is exhausted from the

 

Repair is necessary because of slow

 

exhaust is greater

cuff, and an error code is

 

constant exhaust.

 

than 60 seconds.

displayed.

 

 

 

 

 

 

 

E32

Clock error.

An error code is displayed.

 

If you not clear this error, the product

 

 

 

 

needs repair.

 

 

 

 

 

E50

Pressure offset

An error code is displayed

 

Release the air from the cuff com-

 

error to measure

at restarting the product.

 

pletely, reset the product. If you not

 

pulsation.

 

 

clear this error, repair is necessary.

 

 

 

 

 

E52

Memory error.

An error code is displayed

 

The product needs repair.

 

 

at restarting the product.

 

 

 

 

 

 

 

E53

Battery contact is

The measurement is

 

Replace batteries correctly. If you can

 

defective.

stopped, air is released

 

not clear this error, the product needs

 

 

from the cuff and an error

 

repair.

 

 

code is displayed.

 

 

 

 

 

 

 

E55

Exhaust error.

An error code is displayed

 

Relax and do not move during the

E56

 

at measurement.

 

measurement. If this error occurs

E57

 

 

 

many times, repair is necessary.

 

 

 

 

 

E60

Interval setting

Start time is not proper,

 

Enter parameters for the interval

 

error.

interval of last block is not

 

correctly.

 

 

set in the unit of 120 min.

 

 

 

 

 

 

 

E70

RS-232C error.

The error code is displayed

 

Re-connect the communication cable.

E71

 

during communications.

 

If you can not clear this error, the

E72

 

 

 

product needs repair.

E73

 

 

 

 

 

 

 

 

 

E74

Low battery for

 

 

Replace batteries with new ones and

 

communication.

 

 

restart communication.

 

 

 

 

 

E75

Protcol error due

 

 

Re-connect the communication cable.

 

to external

 

 

If you can not clear this error, the

 

equipment.

 

 

product needs repair.

 

 

 

 

 

E90

Pressure zero

This error code is

 

Release the air from the cuff

 

error for safety

displayed before the

 

completely.

 

circuit.

measurement.

 

 

 

 

 

 

 

E91

Safety circuit

Patient moved during the

 

Relax and try to quiet during the

 

detects over load

measurement.

 

measurement. If it occurs in quiet, the

 

pressure.

 

 

recorder needs repair.

 

 

 

 

 

Other

 

Monitor code is displayed.

 

Reset. Turn on power switch again.

 

 

 

 

 

Maintenance

Page 33

Error Code

Image 35
Contents TM-2430 Page Contents Definitions Before useCompliance Precautions for Use During use Periodic maintenanceStorage Before usePacking List and Component Names WelcomeProduct overview Welcome and IntroductionAir socket Display SymbolsFunctions and Specifications SpecificationsFeatures Pressurization Automatic measurementStopping a measurement Concealing the measurement valuePerformance specifications Environment specifications Steps for replacing the batteries Activating the recorderReplacing Batteries Display & key Conditions When Recorder Switched OnWhen recorder is Condition Actions Turned on Parameters for the Display and ClockSteps for setting the display and clock Mode Steps for selecting a modeSelection for the Automatic Measurement Set up procedure Mode III SettingsAs the start time of the third block Steps for automatic measurementFrom previous Steps for deleting old data Resetting the RecorderSteps for reset Deleting Old DataManual measurement Stopping or canceling an automatic measurementPreparing the patient Patient instructionsSteps for attaching the cuff and recorder Use of the cuff coverAttaching the cuff and recorder Range, you need a proper cuff Preparation of the Carrying Case Using the BeltUsing the Shoulder Band To stop a Current Measurement OperationAutomatic Measurement Manual MeasurementData Transmission to a Printer Specifications for an adaptable printerData Transfer Print sample Steps for data transmissionData Transmission to a Computer Using Analysis Soft Analysis Software and Communication Cables Options and accessoriesOthers Cuffs and Other AccessoriesCuff sleeves Cuff cover Steps for checking accuracy Maintenance Checking Accuracy Required equipmentPeriodical inspection Cleaning the cuff and recorderProblem solving Error code updates without announcement Error codesMaintenance Error Code Memo