A&D Ambulatory Blood Pressure Monitor Operation, Automatic Measurement, Manual Measurement

Page 26

Operation

Automatic Measurement

Caution

Automatic measurement uses the internal clock and parameters of automatic measurement. Refer to section "Parameters for the Display and Clock" and "Selection for the automatic measurement" for setting these parameters.

Pressandholdthe AUTO ON/OFF keyforapprox.3secondssoastoturnoff"A"symbol on display, when the patient stops the automatic measurement or detaches the cuff. If the automatic measurement mode is running without a patient arm or resistance, damage will occur to the cuff.

Starting or re-starting automatic measurement mode

Step 1

Step 2

Confirm the parameters for automatic measurement. Refer to "Selection for the automatic measurement".

Press and hold the AUTO ON/OFF key for about 3 seconds. When the "A" is displayed the recorder starts an automatic measurement based on the internal clock and the parameters for automatic measurement.

Operation using mode II (Sleep mode)

Step

1

Press the

 

 

Press the

 

 

Step

2

 

 

 

AUTO ON/OFF

AUTO ON/OFF

key turning off the "S" when the patient wakes up.

key turning on the "S" when the patient goes to bed.

Stopping or canceling automatic measurement

Step 1 When the AUTO ON/OFF key is pressed for about 3 seconds the recorder exits the automatic measurement mode and the "A" disappears. Conversly, by pressing the AUTO ON/OFF key again the "A" will reappear.

Manual Measurement

Step 1 Press the START STOP key. The recorder starts a measurement at once. The results are displayed and stored in memory.

To stop a Current Measurement

 

 

START STOP

 

 

Step 1

Press the

key during measurement. The recorder will stop the

 

measurement at once and releases the air from the cuff.

 

 

 

Automatic Measurement

Page 24

Operation

Image 26
Contents TM-2430 Page Contents Definitions Before useCompliance Precautions for Use Before use Periodic maintenanceStorage During useWelcome and Introduction WelcomeProduct overview Packing List and Component NamesAir socket Symbols DisplayFunctions and Specifications SpecificationsFeatures Concealing the measurement value Automatic measurementStopping a measurement PressurizationPerformance specifications Environment specifications Steps for replacing the batteries Activating the recorderReplacing Batteries Parameters for the Display and Clock Conditions When Recorder Switched OnWhen recorder is Condition Actions Turned on Display & keySteps for setting the display and clock Mode Steps for selecting a modeSelection for the Automatic Measurement Mode III Settings Set up procedureSteps for automatic measurement As the start time of the third blockFrom previous Deleting Old Data Resetting the RecorderSteps for reset Steps for deleting old dataPatient instructions Stopping or canceling an automatic measurementPreparing the patient Manual measurement Steps for attaching the cuff and recorder Use of the cuff cover Attaching the cuff and recorder Range, you need a proper cuff Preparation of the Carrying Case Using the BeltUsing the Shoulder Band Manual Measurement OperationAutomatic Measurement To stop a Current MeasurementData Transmission to a Printer Specifications for an adaptable printerData Transfer Steps for data transmission Print sampleData Transmission to a Computer Using Analysis Soft Options and accessories Analysis Software and Communication CablesCuff cover Cuffs and Other AccessoriesCuff sleeves OthersRequired equipment MaintenanceChecking Accuracy Steps for checking accuracyPeriodical inspection Cleaning the cuff and recorderProblem solving Error codes Error code updates without announcementMaintenance Error Code Memo