A&D Ambulatory Blood Pressure Monitor Selection for the Automatic Measurement, Mode

Page 17

Selection for the Automatic Measurement

This setting initializes measurement intervals that is based on the internal 24- hour clock.

Mode

 

 

 

mode I

07:00 ~ 21:59 The measurement is performed every quarter hour.

 

22:00 ~ 06:59 The measurement is performed every half hour.

 

 

 

 

mode II

The

AUTO ON/OFF

key is pressed at rising and going to bed so that the

 

measurement intervals are changed and the time during sleep can be

 

distinguished on the data.

 

When the "S" is off, the measurement is performed every quarter hour.

 

When the "S" is displayed, the measurement is performed every half

 

hour.

mode III

The measurement interval can change six times within a maximum of 24-

 

hours. ( The recorder can store six measurement intervals (blocks) in 24-

 

hours. A block consists of a start time and frequency.)

Display & key

Mode

START

STOP

Sign of setting parameters for automatic measuremnt

AUTO

ON/OFF

Selecting a mode

Saving a mode

Steps for selecting a mode

ex. Mode II is selected.

Step 1

Press and hold the

START STOP

key for about 3 seconds. The

 

current mode is displayed.

 

 

 

 

 

 

 

 

 

 

 

Press the

AUTO ON/OFF

key so as to display

of

Step 2

 

mode II.

 

 

 

 

 

 

 

 

Press the

START STOP

key. The recorder stores the mode and

Step 3

 

displays the clock.

 

 

 

START STOP

AUTO ON/OFF

START STOP

Initializing the recorder

Page 15

Selection for the Automatic Measurement

Image 17
Contents TM-2430 Page Contents Definitions Before useCompliance Precautions for Use Storage Periodic maintenanceBefore use During useProduct overview WelcomeWelcome and Introduction Packing List and Component NamesAir socket Display SymbolsFunctions and Specifications SpecificationsFeatures Stopping a measurement Automatic measurementConcealing the measurement value PressurizationPerformance specifications Environment specifications Steps for replacing the batteries Activating the recorder Replacing Batteries When recorder is Condition Actions Turned on Conditions When Recorder Switched OnParameters for the Display and Clock Display & keySteps for setting the display and clock Mode Steps for selecting a modeSelection for the Automatic Measurement Set up procedure Mode III SettingsAs the start time of the third block Steps for automatic measurementFrom previous Steps for reset Resetting the RecorderDeleting Old Data Steps for deleting old dataPreparing the patient Stopping or canceling an automatic measurementPatient instructions Manual measurementSteps for attaching the cuff and recorder Use of the cuff coverAttaching the cuff and recorder Range, you need a proper cuff Preparation of the Carrying Case Using the BeltUsing the Shoulder Band Automatic Measurement OperationManual Measurement To stop a Current MeasurementData Transmission to a Printer Specifications for an adaptable printerData Transfer Print sample Steps for data transmissionData Transmission to a Computer Using Analysis Soft Analysis Software and Communication Cables Options and accessoriesCuff sleeves Cuffs and Other AccessoriesCuff cover OthersChecking Accuracy MaintenanceRequired equipment Steps for checking accuracyPeriodical inspection Cleaning the cuff and recorderProblem solving Error code updates without announcement Error codesMaintenance Error Code Memo