Maintenance
Checking Accuracy
Required equipment
❏Accurate office mercury sphygmomanometer or aneroid gauge with inflation system.
❏
❏A rigid cylinder sized to fit the cuff pressured.
Steps for checking accuracy
Step 1 Turn off the
Step 2 Construct the check system as this drawing.
(not included)
Recorder |
| START STOP switch |
| ||||||||
|
|
|
|
|
|
|
|
|
|
| |
|
|
|
|
|
|
|
|
| PO | TS |
|
|
|
|
|
|
|
|
|
| TAS |
| |
|
|
|
|
|
|
|
|
| TR |
| |
|
|
|
|
|
|
|
|
|
|
| |
|
|
|
|
|
|
|
| O/ | NO |
|
|
|
|
|
|
|
|
|
| FF | A |
| Cuff |
|
|
|
|
|
|
|
| TOU |
|
| |
|
|
|
| MIT/ESLAUID |
|
| |||||
|
|
|
|
|
|
| P |
|
|
| |
R I | A |
| E |
|
| NI/ | SY | S |
|
|
|
|
| VAR | ET |
|
|
|
| ||||
|
|
|
|
|
|
|
| ||||
|
| L |
|
|
|
|
|
| |||
|
|
|
|
|
|
|
|
|
|
| |
Power switch |
|
|
|
|
|
|
|
|
|
| A rigid cylinder |
|
| ||||||||||
Diagnostic Kit |
|
|
|
|
|
|
|
|
|
|
Mercury sphyg- | |
momanometer | |
Inflator bulb | 300 |
| 280 |
| 260 |
| 240 |
| 220 |
| 200 |
| 180 |
| 160 |
| 140 |
Exhaust valve and | 120 |
100 | |
adjustment screw | 80 |
60 | |
| 40 |
| 20 |
| 0 |
Step 4 |
Turn on the power switch, when you press and hold the
The
START STOP
key.
Step 5 Squeeze the inflator bulb until cuff pressure reaches to 50 mmHg.
Verify that the difference between the blinking display of
Step 6 Squeeze the inflator bulb until cuff pressure reaches to 150 mmHg. Verify that this difference is within ±3 mmHg.
Step 7 Squeeze the inflator bulb until cuff pressure reaches to 250 mmHg. Verify that this difference is within ±3 mmHg.
Step 8 Release the cuff air, turn off the
This blood pressure recorder is a precision instrument. Contact your nearest
A&D office for this inspection, if you need repair.
Checking Accuracy | Page 30 | Maintenance |