A&D Ambulatory Blood Pressure Monitor Maintenance, Checking Accuracy, Required equipment

Page 32

Maintenance

Checking Accuracy

Required equipment

Accurate office mercury sphygmomanometer or aneroid gauge with inflation system.

TM-2430 Accuracy Diagnostic Kit (TM2430-90).

A rigid cylinder sized to fit the cuff pressured.

Steps for checking accuracy

Step 1 Turn off the TM-2430 and remove the air hose from the unit.

Step 2 Construct the check system as this drawing.

(not included)

Recorder

 

START STOP switch

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

PO

TS

 

 

 

 

 

 

 

 

 

 

TAS

 

 

 

 

 

 

 

 

 

 

TR

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

O/

NO

 

 

 

 

 

 

 

 

 

 

FF

A

 

Cuff

 

 

 

 

 

 

 

 

TOU

 

 

 

 

 

 

MIT/ESLAUID

 

 

 

 

 

 

 

 

 

P

 

 

 

R I

A

 

E

 

 

NI/

SY

S

 

 

 

 

 

VAR

ET

 

 

 

 

 

 

 

 

 

 

 

 

 

 

L

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

Power switch

 

 

 

 

 

 

 

 

 

 

A rigid cylinder

TM-2430 Accuracy

 

 

Diagnostic Kit

 

 

 

 

 

 

 

 

 

 

Mercury sphyg-

momanometer

Inflator bulb

300

 

280

 

260

 

240

 

220

 

200

 

180

 

160

 

140

Exhaust valve and

120

100

adjustment screw

80

60

 

40

 

20

 

0

Step 4

Turn on the power switch, when you press and hold the

The TM-2430 blinks the "0" of measurement value.

START STOP

key.

Step 5 Squeeze the inflator bulb until cuff pressure reaches to 50 mmHg.

Verify that the difference between the blinking display of TM-2430 and mercury sphygmomanometer is within ±3 mmHg.

Step 6 Squeeze the inflator bulb until cuff pressure reaches to 150 mmHg. Verify that this difference is within ±3 mmHg.

Step 7 Squeeze the inflator bulb until cuff pressure reaches to 250 mmHg. Verify that this difference is within ±3 mmHg.

Step 8 Release the cuff air, turn off the TM-2430 and remove the kit.

This blood pressure recorder is a precision instrument. Contact your nearest

A&D office for this inspection, if you need repair.

Checking Accuracy

Page 30

Maintenance

Image 32
Contents TM-2430 Page Contents Definitions Before useCompliance Precautions for Use Periodic maintenance StorageBefore use During useWelcome Product overviewWelcome and Introduction Packing List and Component NamesAir socket Symbols DisplayFunctions and Specifications SpecificationsFeatures Automatic measurement Stopping a measurementConcealing the measurement value PressurizationPerformance specifications Environment specifications Steps for replacing the batteries Activating the recorderReplacing Batteries Conditions When Recorder Switched On When recorder is Condition Actions Turned onParameters for the Display and Clock Display & keySteps for setting the display and clock Mode Steps for selecting a modeSelection for the Automatic Measurement Mode III Settings Set up procedureSteps for automatic measurement As the start time of the third blockFrom previous Resetting the Recorder Steps for resetDeleting Old Data Steps for deleting old dataStopping or canceling an automatic measurement Preparing the patientPatient instructions Manual measurementSteps for attaching the cuff and recorder Use of the cuff coverAttaching the cuff and recorder Range, you need a proper cuff Preparation of the Carrying Case Using the BeltUsing the Shoulder Band Operation Automatic MeasurementManual Measurement To stop a Current MeasurementData Transmission to a Printer Specifications for an adaptable printerData Transfer Steps for data transmission Print sample Data Transmission to a Computer Using Analysis Soft Options and accessories Analysis Software and Communication CablesCuffs and Other Accessories Cuff sleevesCuff cover OthersMaintenance Checking AccuracyRequired equipment Steps for checking accuracyPeriodical inspection Cleaning the cuff and recorderProblem solving Error codes Error code updates without announcementMaintenance Error Code Memo