3.Instrument components and specifications

3.3 Technical data General

Approvals:

The instrument-specific marks are located next to

 

the identification label.

Nominal supply voltages:

100 / 120 / 230 / 240 V AC ±10%

Nominal frequency:

50/60 Hz

Power draw:

70 VA

Protective class :

I

Power fuses:

2 x T 1.0 A UL listed

Pollution degree: :

2

Overvoltage installation :

II

Maximum heat emission:

70 J/s

Operating temperature range:

+10°C to +35°C

Temperature range during storage:

+5°C to +55°C

Relative humidity:

max. 80% non-condensing

Humidity during storage:

< 80%

according to IEC-1010, UL 3101, EN 61010

Dimensions and weight

Basic instrument

Width (including handwheel):

413 mm

Width (excluding handwheel):

300 mm

Depth (including waste tray):

618 mm

Height (total):

305 mm (with tray on the hood)

Working height (knife blade):

100 mm (measured from the base plate)

Working height (knife blade):

168 mm (measured from the table)

Weight (without accessories):

approx. 37 kg

Control panel

 

Width:

94 mm

Depth:

164 mm

Height:

50 mm

Height (in inclined position):

81 mm

Weight (net):

approx. 0.450 kg

14

Instruction manual V 1.4 – 02/2007

Image 14
Contents Leica RM2245 Page Leica Microsystems Nussloch GmbH Table of Contents Table of Contents Qualification of personnel Symbols in the text and their meaningsImportant notes Intended use of instrumentSafety Safety notesPersonal safety precautions Safety Integrated safety devices Safety Leica RM2245 Instrument components and specificationsOverview instrument components Instrument specifications Rear side of instrumentControl panel Technical data GeneralDimensions and weight Basic instrument Microtome Startup Standard deliveryInstallation site requirements Setting up the instrumentChecking the voltage Assembling the handwheelElectrical connections 23 24 Switching on the instrument Instrument control panel OperationOperating elements and their functions Control panel Display and control elements How to activate the Step function Sectioning modeTrimming mode Specimen retraction Mum of 100 μm using the buttons on the control panel Display 4 lights upObject head Step function is disabled in the remaining feed rangeIndication of remaining horizontal feed Directional fixture for specimen clamps To release the clamp, turn the eccentric lever 29 forwardsImportant! Never turn the screw more than 1/2 Turn at a time Inserting the knife holder Inserting the universal cassette clamp Adjusting the clearance angle Example Enlarged detail showing a clearance angle settingClamping the specimen Clamping the knife / disposable bladeCutting into the specimen trimming Sectioning Changing the specimen or interrupting sectioning Finishing daily routineRigid fixture for specimen clamps Optional accessoriesAssembly for fixture for specimen clamps Directional fixture for specimen clampsFine-directional fixture for specimen clamps Quick clamping system Vee insert Specimen clamps and holdersStandard specimen clamp Foil clamp type Optional accessories Universal cassette clamp Universal cassette clamp, ice-cooledSuper mega-cassette clamp Holder for round specimens Optional accessories Knife holder E/E-TC To clamp the blade, rotate clamping lever 10 back upwardsOptional accessories Use Knife holder N/NZ Optional accessories Section waste tray BacklightingInstrument TrayFreezer pack Attach baseplate Universal microscope carrierAssembling the universal microscope carrier Attach horizontal arm Magnifying lens Plug the power plug into the power socket Cold light source Fiber-optic light guideOptional accessories Ordering informationFiber-optic light guide Cold light sourcesLeica CLS Instrument malfunctions Customized solutionsError reports Malfunctions, possible causes and troubleshooting Problem Possible cause Corrective actionPossible faults Clean using an alcohol-based solution or acetone Cleaning and maintenanceCleaning the instrument Cleaning and maintenance Maintenance Replacing fusesMaintenance instructions Lubricating the instrument Warranty and service WarrantyDag Graupner Managing Director EC Declaration of ConformityWe herewith declare, in exclusive responsibility, that