8. Cleaning and maintenance

8.1 Cleaning the instrument

Always remove the knife / blade before detaching the knife holder from the instrument. Always put the knives back into the knife case when not in use!

Never place a knife anywhere with the cutting edge facing upwards and never try to catch a falling knife!

When using cleaners, comply with the safety instructions of the manufacturer and the labor-safety regulations of your laboratory!

When cleaning the outer surfaces, do not use xylene, scouring powders or solvents containing acetone or xylene. Xylene or acetone will damage the finished surfaces!

Ensure that liquids do not enter the interior of the instrument during cleaning!

Before each cleaning carry out the following preparatory steps:

Raise the specimen clamp to the upper end position and activate the handwheel lock.

Switch the unit off and unplug it.

Remove the blade from the knife holder and insert it in the receptacle at the bottom of the dispenser, or remove the knife from the knife holder and put it back in the knife case.

Remove knife holder base and knife holder for cleaning.

Remove the specimen from the specimen clamp.

Remove section waste with a dry brush.

Remove specimen clamp and clean separately.

Instrument and outside surfaces

If necessary, the varnished outside surfaces of the control panels can be cleaned with a mild commercial household cleaner or soap water and then be wiped with a cloth.

To remove paraffin residue, xylene substitutes, paraffin oil, or paraffin removers such as "Para Gard" (Polysciences) can be used.

The instrument must be completely dry before it can be used again.

Cleaning the knife

Always wipe the knife from the back of the knife to the cutting edge. NEVER wipe in the opposite direction - risk of injury!

Clean using an alcohol-based solution or acetone.

62

Instruction manual V 1.4 – 02/2007

Image 62
Contents Leica RM2245 Page Leica Microsystems Nussloch GmbH Table of Contents Table of Contents Qualification of personnel Symbols in the text and their meaningsImportant notes Intended use of instrumentSafety Safety notesPersonal safety precautions Safety Integrated safety devices Safety Leica RM2245 Instrument components and specificationsOverview instrument components Instrument specifications Rear side of instrumentControl panel Technical data GeneralDimensions and weight Basic instrument Microtome Startup Standard deliveryInstallation site requirements Setting up the instrumentChecking the voltage Assembling the handwheelElectrical connections 23 24 Switching on the instrument Instrument control panel OperationOperating elements and their functions Control panel Display and control elements How to activate the Step function Sectioning modeTrimming mode Specimen retraction Mum of 100 μm using the buttons on the control panel Display 4 lights upObject head Step function is disabled in the remaining feed rangeIndication of remaining horizontal feed Directional fixture for specimen clamps To release the clamp, turn the eccentric lever 29 forwardsImportant! Never turn the screw more than 1/2 Turn at a time Inserting the knife holder Inserting the universal cassette clamp Adjusting the clearance angle Example Enlarged detail showing a clearance angle settingClamping the specimen Clamping the knife / disposable bladeCutting into the specimen trimming Sectioning Changing the specimen or interrupting sectioning Finishing daily routineRigid fixture for specimen clamps Optional accessoriesAssembly for fixture for specimen clamps Directional fixture for specimen clampsFine-directional fixture for specimen clamps Quick clamping system Vee insert Specimen clamps and holdersStandard specimen clamp Foil clamp type Optional accessories Universal cassette clamp Universal cassette clamp, ice-cooledSuper mega-cassette clamp Holder for round specimens Optional accessories Knife holder E/E-TC To clamp the blade, rotate clamping lever 10 back upwardsOptional accessories Use Knife holder N/NZ Optional accessories Section waste tray BacklightingInstrument TrayFreezer pack Attach baseplate Universal microscope carrierAssembling the universal microscope carrier Attach horizontal arm Magnifying lens Plug the power plug into the power socket Cold light source Fiber-optic light guideOptional accessories Ordering informationFiber-optic light guide Cold light sourcesLeica CLS Instrument malfunctions Customized solutionsError reports Malfunctions, possible causes and troubleshooting Problem Possible cause Corrective actionPossible faults Clean using an alcohol-based solution or acetone Cleaning and maintenanceCleaning the instrument Cleaning and maintenance Maintenance Replacing fusesMaintenance instructions Lubricating the instrument Warranty and service WarrantyDag Graupner Managing Director EC Declaration of ConformityWe herewith declare, in exclusive responsibility, that