Leica RM2245 instruction manual Attach horizontal arm

Page 54

6.Optional accessories

b

c

Fig. 52

2

A

Fig. 53

2d

hg

e

f

Fig. 54

2. Attach vertical column

Insert the cap screw (b) into the hole of the baseplate from below. Place the lock washer (c) on the cap screw from above.

Thread the silver vertical column (2) onto the baseplate from above and tighten with the Allen key size 8.

The lock washer must be positioned between the baseplate and vertical column to prevent unintentional turning of the column.

Extreme danger of tipping! After installing theverticalcolumn,immediatelyplacethe microtome onto the baseplate so that the front feet of the microtome are positioned in the shallow recesses (A).

3. Attach horizontal arm

Slide the support ring (e) onto the vertical column and position it so that the lock nut (f) is facing to the back. Tighten the lock nut.

Slide the cross-member (d) onto the column. Ensure that the lock screw (g) is facing to the right of the baseplate. The horizontal arm must be centered over the microscope.

Slide the horizontal arm (3), flat side facing the lock screw (g), into the cross-member (d) and tighten.

For more information on connecting and using the microscope, magnifying lens or cold-light source, please see the appropriate manual.

54

Instruction manual V 1.4 – 02/2007

Image 54
Contents Leica RM2245 Page Leica Microsystems Nussloch GmbH Table of Contents Table of Contents Qualification of personnel Symbols in the text and their meaningsImportant notes Intended use of instrumentSafety Safety notesPersonal safety precautions Safety Integrated safety devices Safety Instrument components and specifications Overview instrument componentsLeica RM2245 Instrument specifications Rear side of instrumentTechnical data General Dimensions and weight Basic instrumentControl panel Microtome Startup Standard deliveryInstallation site requirements Setting up the instrumentAssembling the handwheel Electrical connectionsChecking the voltage 23 24 Switching on the instrument Operation Operating elements and their functionsInstrument control panel Control panel Display and control elements Sectioning mode Trimming modeHow to activate the Step function Specimen retraction Mum of 100 μm using the buttons on the control panel Display 4 lights upStep function is disabled in the remaining feed range Indication of remaining horizontal feedObject head Directional fixture for specimen clamps To release the clamp, turn the eccentric lever 29 forwardsImportant! Never turn the screw more than 1/2 Turn at a time Inserting the knife holder Inserting the universal cassette clamp Adjusting the clearance angle Example Enlarged detail showing a clearance angle settingClamping the specimen Clamping the knife / disposable bladeCutting into the specimen trimming Sectioning Changing the specimen or interrupting sectioning Finishing daily routineRigid fixture for specimen clamps Optional accessoriesAssembly for fixture for specimen clamps Directional fixture for specimen clampsFine-directional fixture for specimen clamps Quick clamping system Specimen clamps and holders Standard specimen clampVee insert Foil clamp type Optional accessories Universal cassette clamp Universal cassette clamp, ice-cooledSuper mega-cassette clamp Holder for round specimens Optional accessories Knife holder E/E-TC To clamp the blade, rotate clamping lever 10 back upwardsOptional accessories Use Knife holder N/NZ Optional accessories Section waste tray BacklightingTray Freezer packInstrument Universal microscope carrier Assembling the universal microscope carrierAttach baseplate Attach horizontal arm Magnifying lens Plug the power plug into the power socket Cold light source Fiber-optic light guideOptional accessories Ordering informationCold light sources Leica CLSFiber-optic light guide Customized solutions Error reportsInstrument malfunctions Malfunctions, possible causes and troubleshooting Problem Possible cause Corrective actionPossible faults Cleaning and maintenance Cleaning the instrumentClean using an alcohol-based solution or acetone Cleaning and maintenance Maintenance Replacing fusesMaintenance instructions Lubricating the instrument Warranty and service WarrantyEC Declaration of Conformity We herewith declare, in exclusive responsibility, thatDag Graupner Managing Director