Leica RM2245 instruction manual Important! Never turn the screw more than 1/2 Turn at a time

Page 29

5. Operation

33

Fig. 18

Important! Never turn the screw more than 1/2

turn at a time.

34

Fine adjustment of the force balance

If another accessory of a different weight is mounted on the object head (33), you must check whether it is necessary to readjust the force balance.

Checking the correct setting:

Attach the new accessory and clamp the specimen.

Set the object head to half the height of the vertical travel range by turning the

handwheel (Fig. 18).

If the object head remains in this exact position, the setting is correct.

If the object head moves, i.e. it is raised or low- ered, fine adjustment is necessary.

Failure to adjust the force balance may result in injury while working.

The force balance is adjusted using the screw (34), which can be accessed by removing the section waste tray on the bottom of the base plate of the microtome. Use the Allen key pro- vided, size 5 (with handle!) for the adjustment.

If the object head moves downwards, turn the screw approx. 1/2 turn clockwise.

If the object head moves upwards, turn the screw (34) approx. 1/2 turn counterclockwise.

Continue this procedure until the object head no longer moves once released.

Fig. 19

Leica RM2245

29

Image 29
Contents Leica RM2245 Page Leica Microsystems Nussloch GmbH Table of Contents Table of Contents Important notes Symbols in the text and their meaningsQualification of personnel Intended use of instrumentSafety notes SafetyPersonal safety precautions Safety Integrated safety devices Safety Leica RM2245 Instrument components and specificationsOverview instrument components Rear side of instrument Instrument specificationsControl panel Technical data GeneralDimensions and weight Basic instrument Microtome Standard delivery StartupSetting up the instrument Installation site requirementsChecking the voltage Assembling the handwheelElectrical connections 23 24 Switching on the instrument Instrument control panel OperationOperating elements and their functions Control panel Display and control elements How to activate the Step function Sectioning modeTrimming mode Specimen retraction Display 4 lights up Mum of 100 μm using the buttons on the control panelObject head Step function is disabled in the remaining feed rangeIndication of remaining horizontal feed To release the clamp, turn the eccentric lever 29 forwards Directional fixture for specimen clampsImportant! Never turn the screw more than 1/2 Turn at a time Inserting the knife holder Inserting the universal cassette clamp Example Enlarged detail showing a clearance angle setting Adjusting the clearance angleClamping the knife / disposable blade Clamping the specimenCutting into the specimen trimming Sectioning Finishing daily routine Changing the specimen or interrupting sectioningAssembly for fixture for specimen clamps Optional accessoriesRigid fixture for specimen clamps Directional fixture for specimen clampsFine-directional fixture for specimen clamps Quick clamping system Vee insert Specimen clamps and holdersStandard specimen clamp Foil clamp type Optional accessories Universal cassette clamp, ice-cooled Universal cassette clampSuper mega-cassette clamp Holder for round specimens Optional accessories To clamp the blade, rotate clamping lever 10 back upwards Knife holder E/E-TCOptional accessories Use Knife holder N/NZ Optional accessories Backlighting Section waste trayInstrument TrayFreezer pack Attach baseplate Universal microscope carrierAssembling the universal microscope carrier Attach horizontal arm Magnifying lens Cold light source Fiber-optic light guide Plug the power plug into the power socketOrdering information Optional accessoriesFiber-optic light guide Cold light sourcesLeica CLS Instrument malfunctions Customized solutionsError reports Problem Possible cause Corrective action Malfunctions, possible causes and troubleshootingPossible faults Clean using an alcohol-based solution or acetone Cleaning and maintenanceCleaning the instrument Cleaning and maintenance Replacing fuses MaintenanceMaintenance instructions Lubricating the instrument Warranty Warranty and serviceDag Graupner Managing Director EC Declaration of ConformityWe herewith declare, in exclusive responsibility, that