Leica RM2245 Changing the specimen or interrupting sectioning, Finishing daily routine

Page 35

5. Operation

5.9 Changing the specimen or interrupting sectioning

Lock the handwheel and cover the knife edge with the knife guard prior to any manipulation of knife or object head, as well as prior to changing the specimen block and during all work breaks!

Raise the specimen to the upper end position and activate the me- chanical handwheel lock.

Cover the sectioning edge with the knife guard.

Remove the specimen from the specimen clamp and mount a new sample to continue.

Before cutting into a new specimen, move the object head back to the

rear end position.

5.10Finishing daily routine

Raise the specimen to the upper end position by turning the handwheel and engage the handwheel lock.

Always remove the knife / blade before detaching the knife holder from the instrument.

Always put the knives back into the knife case when not in use! Never place a knife anywhere with the cutting edge facing upwards and never try to catch a falling knife!

Remove the blade from the knife holder and insert it in the receptacle at the bottom of the dispenser, or remove the knife from the knife holder and put it back in the knife case.

Remove the specimen from the specimen clamp.

Push all section debris into the section waste tray and empty the tray.

Switch the instrument off at the main power switch.

Clean the instrument (see Chapter 8.1).

Leica RM2245

35

Image 35
Contents Leica RM2245 Page Leica Microsystems Nussloch GmbH Table of Contents Table of Contents Intended use of instrument Symbols in the text and their meaningsImportant notes Qualification of personnelSafety notes SafetyPersonal safety precautions Safety Integrated safety devices Safety Leica RM2245 Instrument components and specificationsOverview instrument components Rear side of instrument Instrument specificationsControl panel Technical data GeneralDimensions and weight Basic instrument Microtome Standard delivery StartupSetting up the instrument Installation site requirementsChecking the voltage Assembling the handwheelElectrical connections 23 24 Switching on the instrument Instrument control panel OperationOperating elements and their functions Control panel Display and control elements How to activate the Step function Sectioning modeTrimming mode Specimen retraction Display 4 lights up Mum of 100 μm using the buttons on the control panelObject head Step function is disabled in the remaining feed rangeIndication of remaining horizontal feed To release the clamp, turn the eccentric lever 29 forwards Directional fixture for specimen clampsImportant! Never turn the screw more than 1/2 Turn at a time Inserting the knife holder Inserting the universal cassette clamp Example Enlarged detail showing a clearance angle setting Adjusting the clearance angleClamping the knife / disposable blade Clamping the specimenCutting into the specimen trimming Sectioning Finishing daily routine Changing the specimen or interrupting sectioningDirectional fixture for specimen clamps Optional accessoriesAssembly for fixture for specimen clamps Rigid fixture for specimen clampsFine-directional fixture for specimen clamps Quick clamping system Vee insert Specimen clamps and holdersStandard specimen clamp Foil clamp type Optional accessories Universal cassette clamp, ice-cooled Universal cassette clampSuper mega-cassette clamp Holder for round specimens Optional accessories To clamp the blade, rotate clamping lever 10 back upwards Knife holder E/E-TCOptional accessories Use Knife holder N/NZ Optional accessories Backlighting Section waste trayInstrument TrayFreezer pack Attach baseplate Universal microscope carrierAssembling the universal microscope carrier Attach horizontal arm Magnifying lens Cold light source Fiber-optic light guide Plug the power plug into the power socketOrdering information Optional accessoriesFiber-optic light guide Cold light sourcesLeica CLS Instrument malfunctions Customized solutionsError reports Problem Possible cause Corrective action Malfunctions, possible causes and troubleshootingPossible faults Clean using an alcohol-based solution or acetone Cleaning and maintenanceCleaning the instrument Cleaning and maintenance Replacing fuses MaintenanceMaintenance instructions Lubricating the instrument Warranty Warranty and serviceDag Graupner Managing Director EC Declaration of ConformityWe herewith declare, in exclusive responsibility, that