Leica RM2245 instruction manual Holder for round specimens

Page 44

6.Optional accessories

6.2.7 Holder for round specimens

The holder for round specimens is designed to accommodate cylindrical samples. Inserts for specimens of 6, 15 and 25 mm diameter are available.

89.1

89.2

92

91

• To mount the required insert (89.1-3) turn the

clamping ring (90) counterclockwise and

remove it.

• Place the required insert into the tension

ring (90) and screw the tension ring onto the

thread (91) by turning it clockwise.

• Mount the sample (67) and fix by turning the

clamping ring (90) clockwise.

• To orient the inserted sample, insert the pin

89.3

90

67

93

(92) into the bore (93) and rotate it counter-

clockwise to release the clamp. You can

now rotate the specimen so that the side

you want faces upwards.

• To lock it in the position you have chosen,

tighten the pin (92) by turning it clockwise.

Fig. 36

44

Instruction manual V 1.4 – 02/2007

Image 44
Contents Leica RM2245 Page Leica Microsystems Nussloch GmbH Table of Contents Table of Contents Symbols in the text and their meanings Important notesQualification of personnel Intended use of instrumentSafety Safety notesPersonal safety precautions Safety Integrated safety devices Safety Leica RM2245 Instrument components and specificationsOverview instrument components Instrument specifications Rear side of instrumentControl panel Technical data GeneralDimensions and weight Basic instrument Microtome Startup Standard deliveryInstallation site requirements Setting up the instrumentChecking the voltage Assembling the handwheelElectrical connections 23 24 Switching on the instrument Instrument control panel OperationOperating elements and their functions Control panel Display and control elements How to activate the Step function Sectioning modeTrimming mode Specimen retraction Mum of 100 μm using the buttons on the control panel Display 4 lights upObject head Step function is disabled in the remaining feed rangeIndication of remaining horizontal feed Directional fixture for specimen clamps To release the clamp, turn the eccentric lever 29 forwardsImportant! Never turn the screw more than 1/2 Turn at a time Inserting the knife holder Inserting the universal cassette clamp Adjusting the clearance angle Example Enlarged detail showing a clearance angle settingClamping the specimen Clamping the knife / disposable bladeCutting into the specimen trimming Sectioning Changing the specimen or interrupting sectioning Finishing daily routineOptional accessories Assembly for fixture for specimen clampsRigid fixture for specimen clamps Directional fixture for specimen clampsFine-directional fixture for specimen clamps Quick clamping system Vee insert Specimen clamps and holdersStandard specimen clamp Foil clamp type Optional accessories Universal cassette clamp Universal cassette clamp, ice-cooledSuper mega-cassette clamp Holder for round specimens Optional accessories Knife holder E/E-TC To clamp the blade, rotate clamping lever 10 back upwardsOptional accessories Use Knife holder N/NZ Optional accessories Section waste tray BacklightingInstrument TrayFreezer pack Attach baseplate Universal microscope carrierAssembling the universal microscope carrier Attach horizontal arm Magnifying lens Plug the power plug into the power socket Cold light source Fiber-optic light guideOptional accessories Ordering informationFiber-optic light guide Cold light sourcesLeica CLS Instrument malfunctions Customized solutionsError reports Malfunctions, possible causes and troubleshooting Problem Possible cause Corrective actionPossible faults Clean using an alcohol-based solution or acetone Cleaning and maintenanceCleaning the instrument Cleaning and maintenance Maintenance Replacing fusesMaintenance instructions Lubricating the instrument Warranty and service WarrantyDag Graupner Managing Director EC Declaration of ConformityWe herewith declare, in exclusive responsibility, that