7. Customized solutions
7.1 Instrument malfunctions
Certain instrument malfunctions appear on the
7.1.1 Error reports
40
41
Indication E1
If this message follows the software version indication after switching on the instrument, a loss of all parameters has occurred.
• Press any key.
All parameters are then set back to the default values.
•Now all settings must be checked to see if the previously entered values are still correct.
The values may have to be reentered.
You can continue with your normal working routine.
If
•Proceed as described above.
•Work normally after checking the settings.
•Have the battery replaced by Leica service as soon as possible.
LEDs (40), (41) in the coarse feed buttons are lit at the same time.
This indicates a defect in the detection of the STOP and HOME positions of the object head.
The problem can only be resolved by customer service.