Troubleshooting Alarms
Attr. | Description | Recommended Action | ||
Code | ||||
|
|
| ||
MSTU | Status of the DICOM protocol | Restart the service. | ||
| component: | If the error persists, it may | ||
| • 0 = No Errors | indicate an installation problem. | ||
| • 1 = Not Started | Contact HP technical support. | ||
|
| |||
| • 2 = Invalid/Missing DICOM |
| ||
|
| Dictionary |
| |
| • | 3 | = Starting... |
|
| • | 4 | = Stopping... |
|
| • 5 = Invalid/Missing Index Tag List |
| ||
|
|
| ||
NADC | Number of ADCs the device is connected | Restore connectivity to an ADC as | ||
| to. |
|
| soon as possible. This may require |
| There should always be at least one ADC. | updating authentication | ||
| If there are no connected ADCs, the server | certificates. | ||
| starts drifting in relative time, which may | If this condition persists, contact | ||
| affect system operations and synchroniza- | HP technical support. | ||
| tion over longer periods. |
| ||
|
|
| ||
NAMS | Number of AMSs the device is connected | Restore connectivity to an AMS as | ||
| to. |
|
| soon as possible. This may require |
| If there are no connected AMSs, the | updating authentication | ||
| device cannot report audit events and | certificates. | ||
| they are queued and unavailable to users | If this condition persists, contact | ||
| until the AMS connection is restored. | HP technical support. | ||
|
|
| ||
NANG | Network Auto Negotiate setting: | Check the network adapter config- | ||
| • | 0 | = Off | uration. The setting must match |
| • | 1 | = On | preferences of your network |
| routers and switches. | |||
| • | 2 | = Unavailable | |
| An incorrect setting can have a | |||
| • | 3 | = Unknown | |
| severe impact on grid | |||
| • | 4 | = Unsupported | performance. |
|
|
|
|
|
183
HP Medical Archive