TROUBLESHOOTING
Fault & Alert Messages
| DISPLAYED MESSAGE | CHECK/REPAIR |
| CONDITION/INDICATES | |
|
| |
|
|
|
| “Temp Probe Open” | • Check the J5 plug/socket connections at the |
| (Fault Condition) | CCB ensure they are mating properly and pro- |
|
| viding good contact. Check the pins inside the |
| The control system has detected an open circuit in | J5 plug/socket for wear or damage. See CCB |
| the temperature sensor. The control system will | illustration and socket identification on pages 42 |
| declare this fault condition if it senses a resistance | - 44. Replace damaged plug connectors/wiring |
| above 56,000 ohms from the temperature sensor. | harness as necessary. |
| Possible Causes: | • Check for pinched or broken wiring between the |
| Immersion Temperature Probe and the J5 plug/ | |
| Plug/socket connection problems | socket connection on the CCB - repair or |
| Wiring connection problems | replace damaged wiring as necessary. See |
| Defective Immersion Temperature Probe | CCB illustration and socket identification on |
|
| pages 42 - 44. |
|
| • Check the resistance of the temperature sensor |
|
| inside the Immersion Temperature Probe - see |
|
| pages 37 - 39. Replace the Immersion Temper- |
|
| ature Probe if measured resistance is above |
|
| 56,000 ohms. |
|
| • Call the toll free technical support phone num- |
|
| ber on the back cover of this manual for further |
|
| assistance if the problem has not been cor- |
|
| rected after performing the procedures outlined |
|
| here. |
|
|
|
| “Temp Probe Short” | • Check for pinched or shorted wires between the |
| (Fault Condition) | J5 plug/socket connection on the CCB and the |
|
| Immersion Temperature Probe - repair or |
| The control system has detected a shorted circuit in | replace damaged wiring as necessary. See |
| the temperature sensor. The control system will | CCB illustration and socket identification on |
| declare this fault condition if it senses a resistance | pages 42 - 44. |
| below 390 ohms from the temperature sensor. | • Check the resistance of the temperature sensor |
| Possible Causes: | inside the Immersion Temperature Probe - see |
| pages 37 - 39. Replace the Immersion Temper- | |
| Shorted wiring | ature Probe if measured resistance is below 390 |
| Defective Immersion Temperature Probe | ohms. |
|
| • Call the toll free technical support phone num- |
|
| ber on the back cover of this manual for further |
|
| assistance if the problem has not been cor- |
|
| rected after performing the procedures outlined |
|
| here. |
|
|
|
|
|
|
Servicing should only be performed by a Qualified Service Agent
68