d.Poisonous and/or Corrosive Compounds – When pumping poisonous, reactive, and/or corro- sive gas, vapors, or chemicals, even proper operation and regeneration will not always ensure that all hazardous materials have been totally removed. If hazardous gas, vapors, chemicals, or combustible mixtures are pumped, sufficient quantities may exist during operation or remain after regeneration to cause severe injury or death. Overheating the pump oil, exposing it to air or reac- tive materials, or overpressurizing it above the normal operating range (approximately 1 mTorr (.133 Pa)) will decompose the oil and possibly make it toxic. This is especially true of back- streamed mechanical pump oils which are more volatile (unstable). Overheating of accidentally introduced or backstreamed mechanical pump oils cannot be protected against by thermal switch- es which are set for diffusion pump oil.

d. Refer to specific instruction manuals for detailed instructions and precautions. Always vent the pump and relief valve to a safe location thus ensuring adequate dilution to safe levels, and take all other action required to meet quality air standards. Always handle pump fluids and hardware with an awareness of the possible deadly hazards involved and the necessity for great care and atten- tion to safety precautions.

d.Diffusion pumps are typically cleaned with acetone or alcohol. Acetone, alcohol, and most other solvents are irritants, narcotics, and depressants, and/or carcinogenic. Their inhalation and inges- tion may produce serious effects. Even absorption through the skin can result in moderate toxicity. Always ensure that cleaning operations are performed in large, well-ventilated rooms. Use of self- contained breathing apparatus may be necessary depending upon the solvent type and vapor concentration in surrounding air.

e.High Voltage – Diffusion pumps operate at voltages high enough to kill through electrical shock. Design equipment utilizing these pumps to prevent personnel contact with high voltages. Securely attach prominent hazard warnings. Personnel should always break the primary circuit to the power supply when direct access to the heater or wiring is required.

f.Hot Surfaces – Boiler temperatures reach 530°F (275°C) which can cause serious burns when touched. Always ensure that surfaces have cooled near room temperature before touching them.

g.Hot Coolant and/or Steam – The water used to cool the pump can reach scalding temperatures. Touching or rupture of the cooling surface can cause serious burns. Water left inside quick cool coils from previous use will turn to steam when the pump is reheated. This steam must be allowed to escape without contacting personnel. Whenever possible, design the water system with inter- lock valves so that power cannot be applied to the pump unless water is flowing in the main cool- ing coils (not quick-cooling coils).

h.Cold Surfaces – Liquid nitrogen traps cooled by liquid nitrogen are commonly used in diffusion pumps. Metal surfaces at liquid nitrogen temperature can cause severe frostbite if contacted by unprotected skin. These surfaces remain cold for some time (at least a half hour) after the liquid nitrogen has evaporated.

vi

Image 10
Contents HS-20 Diffusion Pump Sales and Service Offices Telephone Numbers Warranty Warranty Replacement and AdjustmentPage Page Iii Prohibited Action Explosion-Causing Condition Prohibited Action Result Page Vii Page Page Table of Contents Section Safety Considerations Section I InstallationList of Illustrations List of TablesOperating Specifications See Figure Installation GeneralModel Technical Specifications ASA Outline Drawing HS-20 Diffusion Pump, with ASA Flanges Cleaning a NEW Pump UnpackingInstallation Vacuum Connections Safety Considerations and InterlocksUtility and System Connections Cooling Water Connections See FigureCooling Water Connections Thermal Switches Initial Vacuum Test240/400/430/440/480 V Wiring Diagram HS-20 Diffusion Pump 415 V Wiring Diagram HS-20 Diffusion Pump Adding Pump Fluid Sight Glass Assembly START-UP Procedure OperationShutdown Procedure OperationMaintenance General Periodic InspectionMaintenance CleaningDisassembling and Reassembling the Cold CAP Refer to Figure Do not overtighten the screwExternal Fittings Cold Cap Assembly Disassembling the Jet Assembly Refer to Figure JET AssemblyDisassembling the Lower Jet Assembly Inch Jet Assembly Reassembling the Jet Assembly Maintenance Reassembling the Lower Jet AssemblyHeater Replacement Jet Coupling Tighten clamping nuts to 250 in.-lbs Maintenance Replaceable Parts Maintenance Outgassing Troubleshooting LeakagePoor Pump or System Performance Troubleshooting