Installation
1-6 UTILITY AND SYSTEM CONNECTIONS
1-6-1 Safety Considerations and Interlocks
a.Large pumps are equipped with two thermal switches. Both thermal switches on the
WARNING
Utility failure can cause damage to the equipment, overheating, and possible explosions. Designers of the equipment using diffusion pumps must design their system to protect personnel and property from possi- ble hazards.
b.A small lamp of the same voltage as the diffusion pump heater may be connected across the ter- minals of the thermal switch. The lamp will normally be shunted by the contacts. However, when the contacts open, the lamp will receive full line voltage and give a visual trouble indication.
1-6-2 Vacuum Connections
a.The pump must be installed with the body vertical and plumb. Check that the mating flange on the system (to the pump inlet) is horizontal within plus or minus 1 degree. If this condition is not met, correct the system before installing the pump. The boiler of the pump (especially for large pumps) must be horizontal to prevent uneven fluid level and avoid local overheating where the fluid level would be too shallow.
b.Prepare inlet
c.Using appropriate lifting apparatus, align the bolt holes of the flanges and assemble the bolts. Tighten the bolts evenly to compress the
d.Check fill and drain plugs and the sight glass assembly for tightness. Apply light to medium torque, enough to compress gaskets. Minute leaks through the gaskets may develop with use in this location. They may be detectable with very sensitive mass spectrometer leak detectors.
1-6-3 Cooling Water Connections (See Figure 1-3)
The following method of external connections is recommended.
a.Water connections are designed for 1⁄4 FPT fittings. The cold cap cooling coil and all of the body and foreline cooling should be connected in series. A cooling water supply of 1.5 gpm at 60°F to 80°F (16°C to 27°C) should be adequate. A higher flow than this will not be harmful.