SECTION I
INSTALLATION
1-1 GENERAL
Upon receipt, inspect the container for evidence of damage in shipment. Do not discard any evidence of rough handling. Factory packing provides for maximum protection during shipment. However, the pump should be inspected immediately and any damage should be reported to the carrier without delay.
Before unpacking and installing the pump, the user should thoroughly familiarize himself with the contents of this instruction manual, particularly the physical and operating specifications listed in the following paragraphs, in order to gain a better understanding of the operating principles, limitations, correct application of the diffusion pump, the hazards involved, and what to do and what not to do in order to avoid these hazards.
1-2 OPERATING SPECIFICATIONS (See Figure 1-1)
Table
Table 1-1. HS-20 Diffusion Pump Specifications
Model | Technical Specifications |
Optimum Operating Range | 6 x |
| 8 x |
| 7.8 x |
Maximum Pumping Speed | 17,500 liters/sec for air |
(See Figure | 21,000 liters/sec for helium |
Maximum Throughput | 18 |
at 1 x | 2394 |
| 23.4 |
Maximum Forepressure | No Load: 0.65Torr; 86.5 Pa; 0.85 millibar |
| Full Load: 0.55Torr; 73.2 Pa; 0.72 mbar |
Backstreaming Rate at Pump Inlet | 0.0015 mg/cm2/min |
Power Required (approximately) | 12 KW, 240/380/480V, 50/60 Hz, |
Warmup Time | 45 minutes |
Cooldown Time(using quick cool) | 85 minutes (45 minutes) |
Fluid Charge | 5 U.S. quarts (4.75 liters) |
Cooling Water Requirements | 1.5 gpm at 60°F to 80°F inlet temperatures |
Pressure drop across cooling circuits | 15 psi |
of cold cap and body in series |
|
Backing Pump Size Recommended | 100 cfm for maximum throughput |