Installation
1-4 UNPACKING
Pumps are
WARNING
_____________
When lifting some Varian diffusion pumps, use
a.Remove flange covers, blank plugs, protective plugs from water connections, etc. Be careful not to scratch or otherwise damage or mark the sealing surface
b.Carefully remove the outer shipping container and save it for evidence if damaged. Notify the carrier immediately. Visually inspect the pump for damage that may have occurred during shipping. Discard any internal packing which may have been used to protect the internal components.
c.For
d.Check the internal jet assembly. It should be concentric and firmly seated on the bottom of the pump. Using a flashlight, check to make sure that the ejector nozzle is opposite the foreline (the pump outlet connection). The jet assembly should not rotate because of an indexing pin at the bottom.
1-5 CLEANING A NEW PUMP
a.Remove the cold cap as described in Section III, para.
b.Disassemble the internal jet system from the body of the pump. See Figure
c.Remove fill and drain plugs and the sight glass with its
d.Thoroughly clean all components of the jet assembly and pump casing interior (but not