TW-357 REV E

Control Components

Interface Panel

The KRYOS Interface panel, which the user will interact with, contains the vacuum fluorescent display as well as the number keypad, power button, help button and the soft-key control buttons. (See figure 1.0 below.)

Figure 1.0 KRYOS Interface Panel

Main Control

The “brain” for the control system “talks” to the interface unit and makes all decisions regarding liquid levels, temperatures, valve opening/closing, etc. It is located away from the Interface Panel on the upper outer head. (See figure 2.0 below.)

Main Control

Figure 2.0 KRYOS Main Control

Sensor Assembly

A standard 7+1 thermistor assembly, includes the Freeze-Guard over-fill sensor. Optional 4-thermistor, or 8- thermistor sensor assemblies can be ordered. The 4-thermistor assembly maintains the liquid level between 2 middle sensors. The 8-thermistor assembly maintains the liquid level between the high sensor and the low sensor assigned by the user. The standard 7-thermistor assembly is similar to an 8 sensor assembly in that the user can select the START FILL and STOP FILL positions. The eighth position on this assembly is tied into an inline plumbing thermistor, which detects if the solenoid valve fails to close.

12

Image 12
Contents Labs TW-357 REV E Installation Table of ContentsSafety Precautions for Liquid Nitrogen Extreme Cold Cover Eyes and Exposed SkinKeep Equipment Area Well Ventilated Liquid Nitrogen SystemFreight Damage Precautions Dispose of Waste Liquid Nitrogen SafelyElectrical General Information Text Format NotationLabs Cryostorage Specifications Kryos Specifications Installation Unpacking and InspectionLiquid Nitrogen Supply Connection Repackaging for ShipmentPower Supply Connection Kryos Control Field InstallationOperation Initial FillSensor Assembly Control ComponentsInterface Panel Lid Switch Battery BackupRemote Alarm Solenoid ValveLiquid Phase Storage Operating ParametersAdding an Inventory Control System Thermocouple PositioningRacks are not Square Orientation is Important Inventory Control System Layouts Labs Cryostorage SystemLABS-80K LABS-40K Fenwal 4R59953 Blood Bag Rack Arrangements Fenwal 4R5462 9955 Blood Bag Rack Arrangements Power Control OperationIntroduction Normal Fill CycleMain Display Screen Menu SystemHigh Temperature 4.1 Valve Open Temperature Thermocouple Select Calibrate TemperatureTest Temperature System Help ScreensAlarms and Error Conditions Sensor TypeSystem Alarms Test AlarmsLogging Error Logs Dump LogsTemperature Logs System LogsErase Logs SecurityLid Switch Setup Control By TemperatureTemperature Control Settings Control TemperatureSerial Communications and RS-232 Settings User Access LogsTime and Date Set Control Range Freeze-Guard OptionsValve Monitoring Valve Open DurationControl On Battery Backup Version Solenoid ValveDisplay Brightness About this ControlBattery Life Interface SoftwareMaintenance Strainer Cleaning InstructionsDefrosting Your LAB-Series CryoStorage System Cleaning Your Taylor-Wharton CryoStorage System Making Adjustments to the Labs Kryos Sensor Assembly Default Start Fill / Stop Fill Default Auto Defog Kryos Kryos V9.0 & Kryos V9.0B Level ReadingsOffset Value Units of measureRemoving/Installing the Solenoid Valve Ohms to 18 K ohmsSensor Plug Schematic Pin View SchematicsLabs Plumbing Assembly Plumbing AssemblyTroubleshooting SymptomsController Will Not Turn on Indicates High Liquid LevelTemperature reading 10 to 20 degrees warm Indicates Low LN2 SupplyIndicates Open Sensor Lid Open Alarm QCF Quick Chill Feature Will Not OperateFill Solenoid Cycles On and Off Fill Solenoid Makes Humming NoiseDefog Feature Will Not Operate Power Failure AlarmPush Buttons Will Not Respond Liquid Level Readout is IncorrectFactory Settings Replacement Parts Level Controller Electrical/Mechanical Parts80K/40K/20KTaylor-Wharton Technical Services. Phone 800-898-2657 Fax Harsco Corporation