Taylor TW-357 Troubleshooting, Symptoms, Controller Will Not Turn on, Indicates High Liquid Level

Page 40

TW-357 REV E

TROUBLESHOOTING

Symptoms

The key to troubleshooting your LABS CryoStorage system is to determine which component in the system is the source of the problem. Determine if the problem is occurring in any of the following subsystems: Supply Vessel, Transfer Line, Power Source, Temperature, Level Sensing, Security, Lid Switch, Solenoid Valve, Control Display, Alarm System, Communications. After determining which sub-system is having the problem, isolate the problem further by performing sub-system tests. Once the problem is isolated and defined, it will be easier to solve.

Controller Will Not Turn ON

1.Press POWER button. If display is blank and dark go to next step.

2.Check all connections. Start with jacks at the foot of the controller. Pay special attention to jacks labeled “POWER”. Wall transformer must be plugged into an outlet providing AC voltage between 100 and 130 to deliver AC voltage 21 to 30 to the back of the freezer.

Indicates High Liquid Level

1.Determine actual liquid level using a dipstick. Select MENU, LEVEL SENSING, TEST LEVEL SENSORS. An “L” or “G” will indicate individual thermistor status. “L” meaning that the thermistor is submerged in liquid and “G” indicates that the thermistor is in cold Gas. Determine the pressure, and remaining liquid level in the supply cylinder (15 to 22 psi). A fill solenoid valve that freezes (or sticks) open will typically empty the supply cylinder. Replace solenoid valve if it has failed even once.

2.Liquid level is sensed by thermistors located in a sensor tube. If the sensor tube is blocked or iced at the top, the liquid level in the sensor tube may not rise and fall at the same rate as the liquid level in the freezing chamber. Make sure the sensor tube is not obstructed.

3.The pool of LN2 can become turbulent during a fill. Bubbling and splashing can be amplified by the rack arrangement. The deeper the pool, the more turbulent the surface of the pool will be during a fill. The turbulence of the pool surface can splash on the X-High thermistor and cause a false high alarm. Reduce the splashing.

4.Confirm that sensor assembly is responding to changing liquid level with a dip-test.

Mark the sensor assembly at top of tube to assure re-assembly.

Close liquid supply valve at source.

Remove sensor assembly from sensor tube. DO NOT FORCE. Fill solenoid valve should open and LOW LEVEL alarm should be activated.

Select MENU, LEVEL SENSING, TEST LEVEL SENSORS. An “L” or “G” will indicate individual thermistor status (L = Liquid and G = Gas).

Dip each thermistor in succession into LN2 Observe the controller display noting that each thermistor changes from “G” to “L” as each is submerged. Response time may vary.

Return to the main menu and submerge the START FILL THERMISTOR IN LIQUID. NOTE that the low level alarm ceases, fill solenoid valve is still open. Control is flashing FILLING.

Manually press STOP button and note that the fill solenoid valve closes.

Press FILL button to re-open fill solenoid valve.

Continue to lower the sensor until the STOP FILL thermistor is immersed in the LN2 . The fill should stop after a confirming (CHECK) SPLASH GUARD period.

40

Image 40
Contents Labs TW-357 REV E Installation Table of ContentsSafety Precautions for Liquid Nitrogen Extreme Cold Cover Eyes and Exposed SkinKeep Equipment Area Well Ventilated Liquid Nitrogen SystemDispose of Waste Liquid Nitrogen Safely Freight Damage PrecautionsElectrical General Information Text Format NotationLabs Cryostorage Specifications Kryos Specifications Installation Unpacking and InspectionLiquid Nitrogen Supply Connection Repackaging for ShipmentPower Supply Connection Kryos Control Field InstallationOperation Initial FillControl Components Sensor AssemblyInterface Panel Lid Switch Battery BackupRemote Alarm Solenoid ValveLiquid Phase Storage Operating ParametersAdding an Inventory Control System Thermocouple PositioningRacks are not Square Orientation is Important Inventory Control System Layouts Labs Cryostorage SystemLABS-80K LABS-40K Fenwal 4R59953 Blood Bag Rack Arrangements Fenwal 4R5462 9955 Blood Bag Rack Arrangements Power Control OperationIntroduction Normal Fill CycleMain Display Screen Menu SystemHigh Temperature 4.1 Valve Open Temperature Thermocouple Select Calibrate TemperatureTest Temperature System Help ScreensAlarms and Error Conditions Sensor TypeTest Alarms System AlarmsLogging Error Logs Dump LogsTemperature Logs System LogsErase Logs SecurityLid Switch Setup Control By TemperatureTemperature Control Settings Control TemperatureUser Access Logs Serial Communications and RS-232 SettingsTime and Date Set Control Range Freeze-Guard OptionsValve Monitoring Valve Open DurationControl On Battery Backup Version Solenoid ValveDisplay Brightness About this ControlBattery Life Interface SoftwareStrainer Cleaning Instructions MaintenanceDefrosting Your LAB-Series CryoStorage System Cleaning Your Taylor-Wharton CryoStorage System Making Adjustments to the Labs Kryos Sensor Assembly Default Start Fill / Stop Fill Default Auto Defog Kryos Kryos V9.0 & Kryos V9.0B Level ReadingsOffset Value Units of measureRemoving/Installing the Solenoid Valve Ohms to 18 K ohmsSensor Plug Schematic Pin View SchematicsLabs Plumbing Assembly Plumbing AssemblyTroubleshooting SymptomsController Will Not Turn on Indicates High Liquid LevelIndicates Low LN2 Supply Temperature reading 10 to 20 degrees warmIndicates Open Sensor Lid Open Alarm QCF Quick Chill Feature Will Not OperateFill Solenoid Cycles On and Off Fill Solenoid Makes Humming NoiseDefog Feature Will Not Operate Power Failure AlarmPush Buttons Will Not Respond Liquid Level Readout is IncorrectFactory Settings Replacement Parts Level Controller Electrical/Mechanical Parts80K/40K/20KTaylor-Wharton Technical Services. Phone 800-898-2657 Fax Harsco Corporation