TW-357 REV E

Control Operation

This section of the operating manual is for Taylor-Wharton approved equipment that uses the KRYOS controller.

Introduction

The KRYOS temperature and LN2 level controller is designed for easy operation and uninterrupted, reliable service. This controller will maintain the selected temperature and liquid level range of LN2 in your refrigerator as well as provide audible and visual alarms for any alarm conditions that may occur. An “alarm” is any off-normal condition, such as a sensor short or open circuit, a temperature outside of the limit set by the user, or any other condition that would cause the controller to enter the ALARM mode. “Events” are lid openings and closings, solenoid valve openings and closing, and operation of the controller’s relay for remote alarm indication. “Data” refers to periodic alarm and event logging available at the “Data Output” connection.

The KRYOS controller should require no additional attention to control temperature and maintain liquid level if an adequate supply source of liquid nitrogen is maintained. If your protocol calls for you to “top-off” the cryostorage system at the end of a workday or workweek, press the FILL button. The unit will fill to the upper allowable liquid level and stop automatically. You may choose to manually stop the fill by pressing the STOP button at anytime during the fill.

Normal Fill Cycle

When the refrigerator is filled and the controller is operating, the LOW LEVEL and LOW ALARM sensors are immersed in liquid nitrogen. Their resistance values are interpreted by the controller as “in liquid”. At the same time, the FULL LEVEL and the HIGH ALARM sensors are above the liquid pool, sending the controller an “above liquid” signal. As liquid nitrogen evaporates, the liquid level in the refrigerator drops slowly until the LOW LEVEL sensor is above the liquid and sends a different signal to the controller. After a delay sufficient to ensure the signal, the controller interprets this condition as low liquid and opens the fill solenoid, admitting more liquid nitrogen from the supply source.

The refrigerator will fill slowly. The fill continues until the STOP FILL sensor sends the controller a signal that it is now in liquid. The controller will close the liquid supply solenoid to stop the fill. As liquid evaporates, the display will indicate the liquid is at a normal level as the cycle begins again.

The KRYOS temperature and liquid level controller is designed to support a remote alarm. Connections are provided on the back of the controller chassis.

Power

The control can be turned on and off by pressing the Power button followed by “42”, as instructed on the display panel. The two-step shutdown is a precaution against accidental shutdown. Shutdown can also be password protected to prevent users from turning the system on and off under the security section of this manual.

20

Image 20
Contents Labs TW-357 REV E Installation Table of ContentsSafety Precautions for Liquid Nitrogen Extreme Cold Cover Eyes and Exposed SkinKeep Equipment Area Well Ventilated Liquid Nitrogen SystemElectrical Freight Damage PrecautionsDispose of Waste Liquid Nitrogen Safely General Information Text Format NotationLabs Cryostorage Specifications Kryos Specifications Installation Unpacking and InspectionLiquid Nitrogen Supply Connection Repackaging for ShipmentPower Supply Connection Kryos Control Field InstallationOperation Initial FillInterface Panel Sensor AssemblyControl Components Lid Switch Battery BackupRemote Alarm Solenoid ValveLiquid Phase Storage Operating ParametersAdding an Inventory Control System Thermocouple PositioningRacks are not Square Orientation is Important Inventory Control System Layouts Labs Cryostorage SystemLABS-80K LABS-40K Fenwal 4R59953 Blood Bag Rack Arrangements Fenwal 4R5462 9955 Blood Bag Rack Arrangements Power Control OperationIntroduction Normal Fill CycleMain Display Screen Menu SystemHigh Temperature 4.1 Valve Open Temperature Thermocouple Select Calibrate TemperatureTest Temperature System Help ScreensAlarms and Error Conditions Sensor TypeLogging System AlarmsTest Alarms Error Logs Dump LogsTemperature Logs System LogsErase Logs SecurityLid Switch Setup Control By TemperatureTemperature Control Settings Control TemperatureTime and Date Set Serial Communications and RS-232 SettingsUser Access Logs Control Range Freeze-Guard OptionsValve Monitoring Valve Open DurationControl On Battery Backup Version Solenoid ValveDisplay Brightness About this ControlBattery Life Interface SoftwareDefrosting Your LAB-Series CryoStorage System MaintenanceStrainer Cleaning Instructions Cleaning Your Taylor-Wharton CryoStorage System Making Adjustments to the Labs Kryos Sensor Assembly Default Start Fill / Stop Fill Default Auto Defog Kryos Kryos V9.0 & Kryos V9.0B Level ReadingsOffset Value Units of measureRemoving/Installing the Solenoid Valve Ohms to 18 K ohmsSensor Plug Schematic Pin View SchematicsLabs Plumbing Assembly Plumbing AssemblyTroubleshooting SymptomsController Will Not Turn on Indicates High Liquid LevelIndicates Open Sensor Temperature reading 10 to 20 degrees warmIndicates Low LN2 Supply Lid Open Alarm QCF Quick Chill Feature Will Not OperateFill Solenoid Cycles On and Off Fill Solenoid Makes Humming NoiseDefog Feature Will Not Operate Power Failure AlarmPush Buttons Will Not Respond Liquid Level Readout is IncorrectFactory Settings Replacement Parts Level Controller Electrical/Mechanical Parts80K/40K/20KTaylor-Wharton Technical Services. Phone 800-898-2657 Fax Harsco Corporation