Taylor TW-357 manual Labs Cryostorage Specifications

Page 7

TW-357 REV E

LABS Cryostorage Specifications

Liquid nitrogen at atmospheric pressure weighs 1.78 lb./liter (0.8 kg/liter). All Taylor-Wharton Gas Equipment cryostorage systems are designed to support the full weight of liquid nitrogen and a complete stainless steel or aluminum inventory control system with boxes and specimens.

Dimensions

 

80K

40K

20K

External Operating Height

in

42.5

42.5

42.5

(Top of Step to Lid opening)

mm

1080

1080

1080

Step Height

in

11

11

11

 

mm

279

279

279

Overall Height

in.

60

60

60

(Top of Control Interface)

mm

1524

1524

1524

Usable Height, Internal

in.

30.0

30.0

30.0

 

mm

762.0

762.0

762.0

Outside Diameter

in

59.5

45.0

34.0

 

mm

1511

1143

863.6

Internal Working Diameter

in

55.0

40.5

29.5

 

mm

1397

1029

750

Neck Opening

in

24.5

18.0

13.0

Capacity

mm

622

457

330

 

 

 

 

LN2 Capacity

L

1350

606

407

Power Supply0

VAC

16.5

16.5

16.5

Evaporation Rate1

L/day

15.0

9.0

8

Weight, Empty

lb

1550

920

650

 

kg

703

417

295

Maximum Gross Weight2

lb.

3956

2000

1375

Inventory Control System Specifications

kg

1794

907

624

 

 

 

 

No 5.0 x 5.0 Racks3

 

58

30

14

No. Shelves/Rack

 

13

13

13

No. 3.0 x 3.0 Racks4

 

12

8

4

No. Shelves/Rack

 

13

13

13

Vial Capacity, 2 ml5

 

79300

41600

19500

Blood Bag (Fenwall 4R9951)

 

2100

1204

560

Blood Bag (Fenwall 4R9953)

 

1505

820

400

Blood Bag (Fenwall 4R9955)

 

1050

590

290

Footnotes:

0This is the power supply for the standard battery backup version. The Kryos version (No Battery) uses a 24 VAC power supply. For power outlets other than North American 110 VAC contact Taylor-Wharton Customer Service.

1Evaporation rate is nominal. Actual rate may be affected by the nature of the contents, atmospheric conditions, container history, and manufacturing tolerances.

2Includes the empty weight and total LN2 capacity weight.

35.0 in. x 5.0 in. (127 mm x 127 mm) 100 cell box.

43.0 in. x 3.0 in. (76 mm x 76 mm) 25 cell box.

52 ml vial size; 12.5 mm O.D. internal thread.

7

Image 7
Contents Labs TW-357 REV E Table of Contents InstallationLiquid Nitrogen System Safety Precautions for Liquid NitrogenExtreme Cold Cover Eyes and Exposed Skin Keep Equipment Area Well VentilatedDispose of Waste Liquid Nitrogen Safely Freight Damage PrecautionsElectrical Text Format Notation General InformationLabs Cryostorage Specifications Kryos Specifications Repackaging for Shipment InstallationUnpacking and Inspection Liquid Nitrogen Supply ConnectionKryos Control Field Installation Power Supply ConnectionInitial Fill OperationControl Components Sensor AssemblyInterface Panel Solenoid Valve Lid SwitchBattery Backup Remote AlarmOperating Parameters Liquid Phase StorageThermocouple Positioning Adding an Inventory Control SystemInventory Control System Layouts Labs Cryostorage System Racks are not Square Orientation is ImportantLABS-80K LABS-40K Fenwal 4R59953 Blood Bag Rack Arrangements Fenwal 4R5462 9955 Blood Bag Rack Arrangements Normal Fill Cycle PowerControl Operation IntroductionMenu System Main Display ScreenHigh Temperature 4.1 Valve Open Help Screens Temperature Thermocouple SelectCalibrate Temperature Test Temperature SystemSensor Type Alarms and Error ConditionsTest Alarms System AlarmsLogging Dump Logs Error LogsSecurity Temperature LogsSystem Logs Erase LogsControl Temperature Lid Switch SetupControl By Temperature Temperature Control SettingsUser Access Logs Serial Communications and RS-232 SettingsTime and Date Set Valve Open Duration Control RangeFreeze-Guard Options Valve MonitoringAbout this Control Control On Battery Backup VersionSolenoid Valve Display BrightnessInterface Software Battery LifeStrainer Cleaning Instructions MaintenanceDefrosting Your LAB-Series CryoStorage System Cleaning Your Taylor-Wharton CryoStorage System Making Adjustments to the Labs Kryos Sensor Assembly Units of measure Default Start Fill / Stop Fill Default Auto DefogKryos Kryos V9.0 & Kryos V9.0B Level Readings Offset ValueOhms to 18 K ohms Removing/Installing the Solenoid ValvePin View Schematics Sensor Plug SchematicPlumbing Assembly Labs Plumbing AssemblyIndicates High Liquid Level TroubleshootingSymptoms Controller Will Not Turn onIndicates Low LN2 Supply Temperature reading 10 to 20 degrees warmIndicates Open Sensor Fill Solenoid Makes Humming Noise Lid Open AlarmQCF Quick Chill Feature Will Not Operate Fill Solenoid Cycles On and OffLiquid Level Readout is Incorrect Defog Feature Will Not OperatePower Failure Alarm Push Buttons Will Not RespondFactory Settings Level Controller Electrical/Mechanical Parts80K/40K/20K Replacement PartsTaylor-Wharton Technical Services. Phone 800-898-2657 Fax Harsco Corporation