Verification and Performance Tests
This chapter contains test procedures to check the operation of the Agilent 6610xA Power Modules. The required test equipment is specified in Chapter 1. Sample performance test record sheets are included at the end of the chapter. Instructions are given for performing the tests either from a GPIB controller or the MPS keyboard. Two types of procedures are provided: Operation Verification tests and Performance tests.
Operation Verification | These tests do not check all parameters, but comprise a short procedure to verify that the |
| power module is performing properly. |
Performance | These test all the Performance Specifications (not Supplementary Characteristics) listed in |
| Table |
If you encounter failures or
Note | The power module must pass the selftest at |
| module fails selftest, refer to the overall troubleshooting procedures in Chapter 3. |
|
|
SHOCK HAZARD The tests should only be performed by qualified personnel. During the performance of these tests, hazardous voltages may be present at the output of the module.
Operation Verification Tests
To assure that the module is operating properly, without testing all specified parameters, perform the following test procedures:
1.Perform the
2.Perform the Voltage Programming/Readback Accuracy and the Current Programming/Readback Accuracy Performance Tests in this chapter.
Performance Tests
The following paragraphs provide test procedures for verifying the module’s compliance with the specifications listed in Table