Agilent Technologies 6610XA Constant Current CC Tests, CC Setup, CC Load and Line Regulation

Page 15

unloading transients by triggering on the positive and negative slope.

Figure 2-2. Transient Response Waveform

Constant Current (CC) Tests

CC Setup

Follow the general setup instructions in the Measurement Techniques paragraph and the specific instructions given in the following paragraphs.

Current Programming/Readback Accuracy

This test verifies that the current programming and readback are within specification. The accuracy of the current monitoring resistor is determined by the power coefficient (see Table 1-1).

1.Turn off the module and connect the current monitoring resistor directly across the output and a DVM across the resistor. Refer to the "Current Monitoring Resistor" paragraph.

2.Turn on the module and program the output voltage to 5V and the current to zero.

3.Divide the voltage drop (DVM reading) across the current monitoring resistor by its resistance to convert to amps and record this value as (Io). The actual current (Io) and the reading on the keyboard display should be within the limits specified in the Performance Test Record Tables under CC PROGRAMMING @ 0 AMPS, for the particular model being tested.

4.Program the output voltage to 5V and the current to full-scale (see Table 2-1).

5.Divide the voltage drop (DVM reading) across the current monitoring resistor by its resistance to convert to amps and record this value as (Io). The actual current (Io) and the reading on the keyboard display should be within the limits specified in the Performance Test Record Tables under CC PROGRAMMING @ FULL SCALE, for the particular model being tested.

CC Load and Line Regulation

These tests (CC Load Effect and CC Source Effect given below) are tests of the dc regulation of the power module’s output current. To insure that the values read are not the instantaneous measurement of the ac peaks of the output current ripple, several dc measurements should be made and the average of these readings calculated.

Verification and Performance Tests 15

Image 15
Contents Agilent Part No Microfiche Part No Update June Service GuideCertification Safety Summary Printing History Table of Contents Replacement Parts Diagrams Overview About This ManualManual Part Number Description Revisions and Updates ManualRelated Documents FirmwareSafety Considerations Service Tools and EquipmentElectrostatic Discharge Use1 Service Test Equipment Equipment RequiredCritical Specifications Recommended ModelOperation Verification Tests Performance TestsVerification and Performance Tests Current-Monitoring Resistor Setup for Most TestsMeasurement Techniques ProgrammingVoltage Programming/Readback Accuracy CV SetupConstant Voltage CV Tests Current OvervoltageTransient Recovery Time CV Source EffectCV Noise Pard CC Load and Line Regulation CC SetupConstant Current CC Tests Current Programming/Readback AccuracyCC Source Effect CC Load EffectCC Noise Pard Measurement Uncertainty SpecificationPerformance Test Record for Model Agilent 66101A 8V, 16A Test DescriptionPerformance Test Record for Model Agilent 66102A 20V, 7.5A Actual Verification and Performance Tests Performance Test Record for Model Agilent 66103A 35V, 4.5APerformance Test Record for Model Agilent 66104A 60V, 2.5A 1μV 7mV 230μV 0mV 5mV 15mV 15μA 157μA 4μA Performance Test Record for Model Agilent 66106A 200V, 0.75A Page Troubleshooting Sequence Accessing the FlowchartsTroubleshooting Troubleshooting No Output If You Experienced Other Problems With the ModuleOverall Troubleshooting Sheet 1 Overall Troubleshooting Sheet 2 Overall Troubleshooting Sheet 3 Troubleshooting Microprocessor Circuits Sheet 1 Troubleshooting Microprocessor Circuits Sheet 2 Troubleshooting Error Messages Troubleshooting Overvoltage at Turn-On Sheet 1 Troubleshooting Overvoltage at Turn-On Sheet 2 Troubleshooting No Output Sheet 1 Troubleshooting No Output Sheet 2 Troubleshooting No Output Sheet 3 Troubleshooting No Output Sheet 4 Troubleshooting No Output Sheet 5 Troubleshooting High Output Troubleshooting Will Not Overvoltage Troubleshooting CV Accuracy Troubleshooting CC Accuracy 10. Troubleshooting Downprogrammer 11. Troubleshooting DAC Circuit 12. Troubleshooting Readback Circuits Sheet 1 12. Troubleshooting Readback Circuits Sheet 2 Test Points Built-In Test FunctionsTest Point Test Points Description S3 Test Settings Test Description S3 Setting13. S3 Switch Setting Waveforms Troubleshooting Initialization Eeprom Troubleshooting and Initialization14. Eeprom Initialization Program Sheet 1 Troubleshooting 14. Eeprom Initialization Program Sheet 2 Troubleshooting 14. Eeprom Initialization Program Sheet 3 Troubleshooting 14. Eeprom Initialization Program Sheet 4 Troubleshooting Removing the Fan Disassembly ProceduresRemoving the Module and Module Connector Removing the CoverRemoving the Front Panel Board Removing the Front Panel AssemblyRemoving the Power Board Removing the Bias BoardInstalling the Bias Board Installing the Power BoardInstalling the Front Panel Board and Front Panel Assembly Installing the FanInstalling the Cover Page Theory Of Operation AC Input and Bias SuppliesMicroprocessor Circuits CV/CC DACsBlock Diagram Thoery Of Operation CV Amplifier CC AmplifierOr Gates Pulse-Width Modulator, FETS, and Isolation Transformer Downprogrammer CircuitReadback Multiplexer OV Circuit Page Replacement Parts Replacement PartsModel Model Model DIODE-PWR Rect CHOKE-OUTPUT BEAD-FERRITE CONNECTOR, Hood ASSY-FET, HS Shunt .005 Ohms Shunt .01 Ohms Shunt .035 Ohms R305 R306 R307 R308 R309 R310 R311 R312 R313 R314 R315 Model RES-ZERO Ohms RES 9.31K 1% Replacement Parts All Models 66101A-66103A 66104A-66106A Agilent 6610xA Power Modules Parts List Mechanical Model Output Connector Assembly Parts List Reference Designators Page Schematic Sheets Component Location DiagramsDiagrams Schematic Notes Page Page L305 Page Page Module Front Panel Page Page Page OUT Main Board Component Coordinates Page Backdating Agilent Prefix Serial Number Make Model ChangesAgilent Prefix Serial Number Make Number ChangesBackdating Index IndexSchematic Troubleshooting

6610XA specifications

Agilent Technologies has long been recognized for its innovative contributions to the fields of measurement and testing, and the Agilent 6610XA series of power supplies is no exception. Tailored for demanding applications in electronics and advanced research, the 6610XA series exemplifies cutting-edge technology blended with user-friendly features.

One of the main characteristics of the Agilent 6610XA is its ability to provide a stable, accurate, and reliable power supply across a range of applications. With output voltages that can reach as high as 60V and currents up to 10A, the device is versatile enough to cater to a variety of testing requirements. This power supply is ideal for applications including semiconductor testing, device characterization, and more, where precision and consistency are paramount.

Among the standout features of the Agilent 6610XA is its advanced graphical user interface (GUI), which enhances the overall user experience. The intuitive design allows engineers and technicians to monitor and control voltage and current settings easily, making the process of configuring the device both fast and efficient. Additionally, the device has built-in measurement capabilities that can display real-time voltage and current readings, significantly aiding in troubleshooting and performance evaluation.

Furthermore, the 6610XA incorporates a range of communication interfaces including USB, LAN, and GPIB, making it highly adaptable for integration into various automated test setups. This versatility signifies that the power supply can be seamlessly incorporated into existing laboratory environments, promoting productivity and efficiency.

The series also incorporates intelligent protection mechanisms to ensure both user safety and equipment longevity. Features such as overvoltage protection (OVP) and overcurrent protection (OCP) are designed to prevent accidental overloads, safeguarding both the device under test and the power supply itself.

Another key aspect is the series' capability to perform complex programming tasks with ease. With programming capabilities that enable users to set intricate voltage and current profiles, the device supports advanced applications, including load transient testing and sweep testing. This flexibility makes the Agilent 6610XA a valuable asset for any research and development environment looking to enhance testing efficiency and accuracy.

In conclusion, the Agilent 6610XA series power supply stands out for its combination of precision, user-friendliness, advanced communication capabilities, and safety features. These attributes make it a critical tool for engineers and researchers engaged in the rigorous demands of modern electronics testing and evaluation. With its continued commitment to innovation, Agilent Technologies reinforces its position as a leader in providing high-quality solutions for the measurement and testing industry.