Agilent Technologies 6610XA appendix Eeprom Initialization Program Sheet 3 Troubleshooting

Page 54

1010

NEXT I

 

 

 

1020

!

 

 

 

1030

OUTPUT @Ps;"*CLS"

! Clears power supply registers

1040

!

 

 

 

1050

OUTPUT @Ps;"CAL;STATE ON,"

! Turn on cal mode, "0" passcode

1060

!

 

 

 

1070

GOSUB Ps_error

 

! Error if passcode is not "0"!

1080

IF Err THEN

 

 

1090

OUTPUT @Ps;"*IDN?"

! Get data from model # location

1100

ENTER @Ps;Idn$

 

1110

Model=VAL(Idn$[POS(Idn$,”,”)+1] )

 

1120

ELSE

 

 

 

1130

GOTO Start

 

 

1140

END IF

 

 

 

1150

!

 

 

 

1160

OUTPUT @Ps;"CAL:STATE ON,";Model

! Turn on cal mode, passcode =

1170

 

 

 

! data at model number location

1180

!

 

 

 

1190

GOSUB Ps_error

 

! Error if passcode is not same as

1200

 

 

 

! data at model # location

1210

IF Err THEN

 

 

1220

OUTPUT @Ps;"CAL:STATE ON,";Model$[l,4]

! Turn on cal mode, passcode =

1230

 

 

 

! model #

1240

GOSUB Ps_error

 

 

1250

IF Err THEN

 

 

1260

PRINT "Change pass code to the power supply model # or zero. Restart program."

1270

STOP

 

 

1280

ELSE

 

 

 

1290

GOTO Start

 

 

1300

END IF

 

 

1310

END IF

 

 

 

1320

!

 

 

 

1330

Start:

!

 

 

1340

 

!

 

 

1350

INPUT “Select Initialization (I) or Factory preset replacement (F).”,Sel$

1360

CLEAR SCREEN

 

 

1370

SELECT (UPC$(Sel$))

 

1380

CASE "I"

 

! Select Initialization

1390

GOTO Init_eeprom

 

1400

CASE "F"

 

! Select install new factory data

1410

GOTO Fact_preset

 

1420

CASE ELSE

 

 

1430

BEEP

 

 

 

1440

GOTO Start

 

 

1450

END SELECT

 

 

1460

 

!

 

 

1470

Init_eeprom:

!

 

1480

PRINT “Initializing EEPROM”

 

1490

!

 

 

 

1500

FOR I=1 TO 39

 

 

Figure 3-14. EEPROM Initialization Program (Sheet 3 of 5)

54 Troubleshooting

Image 54
Contents Service Guide Agilent Part No Microfiche Part No Update JuneCertification Safety Summary Printing History Table of Contents Replacement Parts Diagrams About This Manual OverviewFirmware Revisions and Updates ManualRelated Documents Manual Part Number DescriptionSafety Considerations Service Tools and EquipmentElectrostatic Discharge Recommended Model Service Test Equipment Equipment RequiredCritical Specifications Use1Operation Verification Tests Performance TestsVerification and Performance Tests Programming Setup for Most TestsMeasurement Techniques Current-Monitoring ResistorCurrent Overvoltage CV SetupConstant Voltage CV Tests Voltage Programming/Readback AccuracyTransient Recovery Time CV Source EffectCV Noise Pard Current Programming/Readback Accuracy CC SetupConstant Current CC Tests CC Load and Line RegulationCC Load Effect CC Source EffectCC Noise Pard Test Description SpecificationPerformance Test Record for Model Agilent 66101A 8V, 16A Measurement UncertaintyPerformance Test Record for Model Agilent 66102A 20V, 7.5A Performance Test Record for Model Agilent 66103A 35V, 4.5A Actual Verification and Performance TestsPerformance Test Record for Model Agilent 66104A 60V, 2.5A 1μV 7mV 230μV 0mV 5mV 15mV 15μA 157μA 4μA Performance Test Record for Model Agilent 66106A 200V, 0.75A Page Troubleshooting Sequence Accessing the FlowchartsTroubleshooting If You Experienced Other Problems With the Module Troubleshooting No OutputOverall Troubleshooting Sheet 1 Overall Troubleshooting Sheet 2 Overall Troubleshooting Sheet 3 Troubleshooting Microprocessor Circuits Sheet 1 Troubleshooting Microprocessor Circuits Sheet 2 Troubleshooting Error Messages Troubleshooting Overvoltage at Turn-On Sheet 1 Troubleshooting Overvoltage at Turn-On Sheet 2 Troubleshooting No Output Sheet 1 Troubleshooting No Output Sheet 2 Troubleshooting No Output Sheet 3 Troubleshooting No Output Sheet 4 Troubleshooting No Output Sheet 5 Troubleshooting High Output Troubleshooting Will Not Overvoltage Troubleshooting CV Accuracy Troubleshooting CC Accuracy 10. Troubleshooting Downprogrammer 11. Troubleshooting DAC Circuit 12. Troubleshooting Readback Circuits Sheet 1 12. Troubleshooting Readback Circuits Sheet 2 Test Points Built-In Test FunctionsTest Point Test Points Description S3 Setting S3 Test Settings Test Description13. S3 Switch Setting Waveforms Troubleshooting Eeprom Troubleshooting and Initialization Initialization14. Eeprom Initialization Program Sheet 1 Troubleshooting 14. Eeprom Initialization Program Sheet 2 Troubleshooting 14. Eeprom Initialization Program Sheet 3 Troubleshooting 14. Eeprom Initialization Program Sheet 4 Troubleshooting Removing the Cover Disassembly ProceduresRemoving the Module and Module Connector Removing the FanRemoving the Bias Board Removing the Front Panel AssemblyRemoving the Power Board Removing the Front Panel BoardInstalling the Power Board Installing the Bias BoardInstalling the Front Panel Board and Front Panel Assembly Installing the FanInstalling the Cover Page CV/CC DACs AC Input and Bias SuppliesMicroprocessor Circuits Theory Of OperationBlock Diagram Thoery Of Operation CV Amplifier CC AmplifierOr Gates Pulse-Width Modulator, FETS, and Isolation Transformer Downprogrammer CircuitReadback Multiplexer OV Circuit Page Replacement Parts Replacement PartsModel Model Model DIODE-PWR Rect CHOKE-OUTPUT BEAD-FERRITE CONNECTOR, Hood ASSY-FET, HS Shunt .005 Ohms Shunt .01 Ohms Shunt .035 Ohms R305 R306 R307 R308 R309 R310 R311 R312 R313 R314 R315 Model RES-ZERO Ohms RES 9.31K 1% Replacement Parts All Models 66101A-66103A 66104A-66106A Agilent 6610xA Power Modules Parts List Mechanical Model Output Connector Assembly Parts List Reference Designators Page Schematic Sheets Component Location DiagramsDiagrams Schematic Notes Page Page L305 Page Page Module Front Panel Page Page Page OUT Main Board Component Coordinates Page Number Changes Agilent Prefix Serial Number Make Model ChangesAgilent Prefix Serial Number Make BackdatingBackdating Index IndexSchematic Troubleshooting

6610XA specifications

Agilent Technologies has long been recognized for its innovative contributions to the fields of measurement and testing, and the Agilent 6610XA series of power supplies is no exception. Tailored for demanding applications in electronics and advanced research, the 6610XA series exemplifies cutting-edge technology blended with user-friendly features.

One of the main characteristics of the Agilent 6610XA is its ability to provide a stable, accurate, and reliable power supply across a range of applications. With output voltages that can reach as high as 60V and currents up to 10A, the device is versatile enough to cater to a variety of testing requirements. This power supply is ideal for applications including semiconductor testing, device characterization, and more, where precision and consistency are paramount.

Among the standout features of the Agilent 6610XA is its advanced graphical user interface (GUI), which enhances the overall user experience. The intuitive design allows engineers and technicians to monitor and control voltage and current settings easily, making the process of configuring the device both fast and efficient. Additionally, the device has built-in measurement capabilities that can display real-time voltage and current readings, significantly aiding in troubleshooting and performance evaluation.

Furthermore, the 6610XA incorporates a range of communication interfaces including USB, LAN, and GPIB, making it highly adaptable for integration into various automated test setups. This versatility signifies that the power supply can be seamlessly incorporated into existing laboratory environments, promoting productivity and efficiency.

The series also incorporates intelligent protection mechanisms to ensure both user safety and equipment longevity. Features such as overvoltage protection (OVP) and overcurrent protection (OCP) are designed to prevent accidental overloads, safeguarding both the device under test and the power supply itself.

Another key aspect is the series' capability to perform complex programming tasks with ease. With programming capabilities that enable users to set intricate voltage and current profiles, the device supports advanced applications, including load transient testing and sweep testing. This flexibility makes the Agilent 6610XA a valuable asset for any research and development environment looking to enhance testing efficiency and accuracy.

In conclusion, the Agilent 6610XA series power supply stands out for its combination of precision, user-friendliness, advanced communication capabilities, and safety features. These attributes make it a critical tool for engineers and researchers engaged in the rigorous demands of modern electronics testing and evaluation. With its continued commitment to innovation, Agilent Technologies reinforces its position as a leader in providing high-quality solutions for the measurement and testing industry.