Agilent Technologies 6610XA appendix Eeprom Initialization Program Sheet 1 Troubleshooting

Page 52

10! Program to initialize EEPROM or move factory preset data in 66101A,

20! 66102A, 66103A, 66104A, 66105A and 66106A power supply modules.

30! Rev A.00.00 dated May 27, 1993

40!

50DIM Init_data(1:39),Model$[6],Idn$[21],Cal_data$[40]

60INTEGER Addr(1:39),Length(1:39)

70

ASSIGN @Ps TO 70500

! Supply must be at address 70500

80CLEAR SCREEN

90!

100 Eprom_data_addr:

! Data address

110DATA 2,6,10,14,18,19,20,24,28,32

120DATA 36,37,38,42,46,50,55,56,57,58

130DATA 62,66,70,74,,110,l11,112,114,118,119

140DATA l20,121,122,124,125,126,127,128,130

150!

160 Eprom_data_len:

! Data for word length

170DATA 4,4,4,4,1,1,4,4,4,4

180DATA 1,1,4,4,4,4,1,1,1,4

190DATA 4,4,4,4,1,1,2,4,1,1

200DATA 1,1,2,1,1,1,1,1,2

210!

220 Eprom_data_101a:

!

! EEPROM data for 66101A

230DATA 409,668,32.2818,8.19,0,66,0,221.586,33.0533,16.38,0

240DATA 66,4,23.5,19,10,0,255,100,20,66101

250DATA 3277.35,-1.74536,1772.04,-1.50366,0,0,4,66101,20,150

260DATA 20,150,28321,40,20,10,83,83,0

270!

280 Eprom_data_102a: !

! EEPROM data for 66102A

290DATA 164.035,32.3969,20.475,0,66,0,440.367,30.0918,7.678,0

300DATA 66,4,9,5.25,24.0,0,255,100,20,66102

310DATA 1311.8,.248047,3523.12,-28.4202,0,0,4,66102,20,150

320DATA 20,150,28321,40,20,10,83,83,0

330!

340 Eprom_data_103a: !

! EEPROM data for 66103A

350DATA 93.7474,32.2944,35.831,0,66,0,735.971,26.127,4.607,0

360DATA 66,4,4,.93986,5.50002,42,0,255,100,20,66103

370DATA 749.745,-.725342,5885.92,-54.0496,0,0,4,66103,20,150

380DATA 20,150,28321,40,20,10,83,83,0

390!

400 Eprom_data_104a: !

! EEPROM data for 66104A

410DATA 54,32.3969,61.425,0,66,0,1320,30.0918,2.559,0

420DATA 66,4,3,5.25,72.0,0,255,100,20,66104

430DATA 437,.248047,10569,-28.4202,0,0,4,66104,20,150

440DATA 20,150,28321,40,20,10,83,83,0

450!

460 Eprom_data_105a: !

! EEPROM data for 66105A

470DATA 27,32.3969,122.85,0,65,0,2640,30.0918,1.28,0

480DATA 66,4,1.5,5,5.25,144,0,255,100,20,66105

490DATA 219,.248047,21000,-28.4202,0,0,4,66105,20,150

500DATA 20,150,28321,40,20,10,82,83,0

Figure 3-14. EEPROM Initialization Program (Sheet 1 of 5)

52 Troubleshooting

Image 52
Contents Service Guide Agilent Part No Microfiche Part No Update JuneCertification Safety Summary Printing History Table of Contents Replacement Parts Diagrams About This Manual OverviewRevisions and Updates Manual Related DocumentsFirmware Manual Part Number DescriptionService Tools and Equipment Safety ConsiderationsElectrostatic Discharge Service Test Equipment Equipment Required Critical SpecificationsRecommended Model Use1Performance Tests Operation Verification TestsVerification and Performance Tests Setup for Most Tests Measurement TechniquesProgramming Current-Monitoring ResistorCV Setup Constant Voltage CV TestsCurrent Overvoltage Voltage Programming/Readback AccuracyCV Source Effect Transient Recovery TimeCV Noise Pard CC Setup Constant Current CC TestsCurrent Programming/Readback Accuracy CC Load and Line RegulationCC Load Effect CC Source EffectCC Noise Pard Specification Performance Test Record for Model Agilent 66101A 8V, 16ATest Description Measurement UncertaintyPerformance Test Record for Model Agilent 66102A 20V, 7.5A Performance Test Record for Model Agilent 66103A 35V, 4.5A Actual Verification and Performance TestsPerformance Test Record for Model Agilent 66104A 60V, 2.5A 1μV 7mV 230μV 0mV 5mV 15mV 15μA 157μA 4μA Performance Test Record for Model Agilent 66106A 200V, 0.75A Page Accessing the Flowcharts Troubleshooting SequenceTroubleshooting If You Experienced Other Problems With the Module Troubleshooting No OutputOverall Troubleshooting Sheet 1 Overall Troubleshooting Sheet 2 Overall Troubleshooting Sheet 3 Troubleshooting Microprocessor Circuits Sheet 1 Troubleshooting Microprocessor Circuits Sheet 2 Troubleshooting Error Messages Troubleshooting Overvoltage at Turn-On Sheet 1 Troubleshooting Overvoltage at Turn-On Sheet 2 Troubleshooting No Output Sheet 1 Troubleshooting No Output Sheet 2 Troubleshooting No Output Sheet 3 Troubleshooting No Output Sheet 4 Troubleshooting No Output Sheet 5 Troubleshooting High Output Troubleshooting Will Not Overvoltage Troubleshooting CV Accuracy Troubleshooting CC Accuracy 10. Troubleshooting Downprogrammer 11. Troubleshooting DAC Circuit 12. Troubleshooting Readback Circuits Sheet 1 12. Troubleshooting Readback Circuits Sheet 2 Built-In Test Functions Test PointsTest Point Test Points Description S3 Setting S3 Test Settings Test Description13. S3 Switch Setting Waveforms Troubleshooting Eeprom Troubleshooting and Initialization Initialization14. Eeprom Initialization Program Sheet 1 Troubleshooting 14. Eeprom Initialization Program Sheet 2 Troubleshooting 14. Eeprom Initialization Program Sheet 3 Troubleshooting 14. Eeprom Initialization Program Sheet 4 Troubleshooting Disassembly Procedures Removing the Module and Module ConnectorRemoving the Cover Removing the FanRemoving the Front Panel Assembly Removing the Power BoardRemoving the Bias Board Removing the Front Panel BoardInstalling the Power Board Installing the Bias BoardInstalling the Fan Installing the Front Panel Board and Front Panel AssemblyInstalling the Cover Page AC Input and Bias Supplies Microprocessor CircuitsCV/CC DACs Theory Of OperationBlock Diagram Thoery Of Operation CC Amplifier CV AmplifierOr Gates Downprogrammer Circuit Pulse-Width Modulator, FETS, and Isolation TransformerReadback Multiplexer OV Circuit Page Replacement Parts Replacement PartsModel Model Model DIODE-PWR Rect CHOKE-OUTPUT BEAD-FERRITE CONNECTOR, Hood ASSY-FET, HS Shunt .005 Ohms Shunt .01 Ohms Shunt .035 Ohms R305 R306 R307 R308 R309 R310 R311 R312 R313 R314 R315 Model RES-ZERO Ohms RES 9.31K 1% Replacement Parts All Models 66101A-66103A 66104A-66106A Agilent 6610xA Power Modules Parts List Mechanical Model Output Connector Assembly Parts List Reference Designators Page Component Location Diagrams Schematic SheetsDiagrams Schematic Notes Page Page L305 Page Page Module Front Panel Page Page Page OUT Main Board Component Coordinates Page Agilent Prefix Serial Number Make Model Changes Agilent Prefix Serial Number MakeNumber Changes BackdatingBackdating Index IndexSchematic Troubleshooting

6610XA specifications

Agilent Technologies has long been recognized for its innovative contributions to the fields of measurement and testing, and the Agilent 6610XA series of power supplies is no exception. Tailored for demanding applications in electronics and advanced research, the 6610XA series exemplifies cutting-edge technology blended with user-friendly features.

One of the main characteristics of the Agilent 6610XA is its ability to provide a stable, accurate, and reliable power supply across a range of applications. With output voltages that can reach as high as 60V and currents up to 10A, the device is versatile enough to cater to a variety of testing requirements. This power supply is ideal for applications including semiconductor testing, device characterization, and more, where precision and consistency are paramount.

Among the standout features of the Agilent 6610XA is its advanced graphical user interface (GUI), which enhances the overall user experience. The intuitive design allows engineers and technicians to monitor and control voltage and current settings easily, making the process of configuring the device both fast and efficient. Additionally, the device has built-in measurement capabilities that can display real-time voltage and current readings, significantly aiding in troubleshooting and performance evaluation.

Furthermore, the 6610XA incorporates a range of communication interfaces including USB, LAN, and GPIB, making it highly adaptable for integration into various automated test setups. This versatility signifies that the power supply can be seamlessly incorporated into existing laboratory environments, promoting productivity and efficiency.

The series also incorporates intelligent protection mechanisms to ensure both user safety and equipment longevity. Features such as overvoltage protection (OVP) and overcurrent protection (OCP) are designed to prevent accidental overloads, safeguarding both the device under test and the power supply itself.

Another key aspect is the series' capability to perform complex programming tasks with ease. With programming capabilities that enable users to set intricate voltage and current profiles, the device supports advanced applications, including load transient testing and sweep testing. This flexibility makes the Agilent 6610XA a valuable asset for any research and development environment looking to enhance testing efficiency and accuracy.

In conclusion, the Agilent 6610XA series power supply stands out for its combination of precision, user-friendliness, advanced communication capabilities, and safety features. These attributes make it a critical tool for engineers and researchers engaged in the rigorous demands of modern electronics testing and evaluation. With its continued commitment to innovation, Agilent Technologies reinforces its position as a leader in providing high-quality solutions for the measurement and testing industry.