Figure 3-4. Typical Display at Power-On
Self-Test Errors
If the supply fails the
HDW ERR CH "N" 8291 FAILED TIMER FAILED CV DAC CH ’’N"
CCDAC CH "N" OV DAC CH "N" FUSE CH ’’N"
Note: "N" specifies the failed output channel number: 1,2,3, or 4 as applicable.
Figure 3-5. Sample Self-Test Failure Display
Checking Out Your Supply Using Local Control
The following procedures use the display and keys on the front panel to check each of your power supply’s outputs. No test equipment, other than a jumper wire (14 AWG), is required to perform these tests. The tests must be repeated for each output of your particular supply. The checkout consists of voltage, overvoltage, and current tests. It is assumed that power has already been turned on, the supply has passed the
Note | The following procedures are identical for all models and for all outputs. Use the OUTPUT SELECT key |
| to select an output to be tested. If an output fails any of the tests, refer to the troubleshooting section in |
| the Service Manual. |
Getting Started 37