Agilent Technologies 6680A, 6684A, 6683A, 6681A, 6682A service manual Electronic Load

Models: 6682A 6680A 6681A 6683A 6684A

1 156
Download 156 pages 32.3 Kb
Page 11
Image 11

Table 2-1. Test Equipment Required

Type

Required Characteristics

Recommended Model

 

 

 

Digital Voltmeter1

Resolution: 10 nV @ 1V

Agilent 3458A

 

Readout: 8 1/2 digits

 

 

Accuracy: 20 ppm

 

Current Monitor Resistor1

Agilent 6682A, 6683A, Agilent 6684A:

Guildline 9230/300

 

0.001Ω ± 0.04%, 300A, 100W

 

 

Agilent 6680A, 6681A

Burster Type 1280

 

100 ∝Ω ± 0.05%, 1000A

 

DC Power Supply

DC Power Source with current capability

Agilent 6680A

 

equal to UUT

 

Electronic Load

Range: Voltage and current range must

3 each Agilent 6050A, w/3 each

 

exceed that of supply under test. Power:

Agilent 60504B per Agilent 6050A

 

5.4KW minimum

for all units

 

or

 

Resistor Load

5 Kilowatt minimum

 

 

Agilent 6680A = 5.7 milliohms 4375W

 

 

Agilent 6681A = 13.8 milliohms 4640W

 

 

Agilent 6682A = 87.5 milliohms. 5040W

 

 

Agilent 6683A = 200 milliohms 5120W

 

 

Agilent 6684A = 312 milliohms 5120W

 

Oscilloscope

Sensitivity: 1mV

Agilent 54504A

 

Bandwidth Limit: 20MHz

 

 

Probe: 1:1 with RF tip

 

 

 

 

RMS Voltmeter

True RMS Bandwidth: 20MHz

Agilent 3400B

 

Sensitivity: 100 ∝V

 

 

 

 

Current Transformer

0.1Volt per ampere: 1Hz to 20MHz

Pearson Model 411

Variable-Voltage

Power: 3 Phase 24KVA; Range:

Superior Powerstat

Transformer (AC Source)

180-235V 47 - 63Hz; 360- 440V

1156DT-3Y, 0-280V, 50A,

 

47 - 63Hz

24.2 KVA or equivalent .

 

 

 

GPIB Controller2

Full GPIB capabilities

HP Series 200/300

1 Required for Operation Verification Tests.

2 Required for remote testing of 668xA models.

Electronic Load

Many of the test procedures require the use of a variable load capable of dissipating the required power. If a variable resistor is used, switches must be used for connecting, disconnecting, and shorting the load resistor. For most tests, an electronic load (see Table 2-1) is easier to use than a variable resistor. However, an electronic load may not be fast enough for testing transient recovery time or may be too noisy for testing noise (PARD). In these cases, fixed load resistors of suitable power dissipation can be used with minor changes to the test procedures given in this chapter.

12 Verification

Page 11
Image 11
Agilent Technologies 6680A, 6684A, 6683A, 6681A, 6682A service manual Electronic Load