Agilent Technologies 6680A, 6684A, 6683A, 6681A, 6682A service manual 1530

Models: 6682A 6680A 6681A 6683A 6684A

1 156
Download 156 pages 32.3 Kb
Page 66
Image 66

1530

CLEAR SCREEN

 

 

1540

PRINT "This program should ONLY be completed if your power supply”

1550

PRINT "EEPROM has been replaced or a component that will effect"

1560

PRINT "the calibration AND the alignment of voltage, overvoltage"

1570

PRINT "and current is complete AND unit has passed the performance"

1580

PRINT "test.

 

Enter C to continue, any other key to abort.”

1590

INPUT Cont_prog$

 

 

1600

IF (UPC$(Cont_prog$))< >"C" THEN GOTO Cal_off

 

1610

!

 

 

 

1620

CLEAR SCREEN

 

 

1630

PRINT "Transferring calibration data to factory preset locations."

1640

!

 

 

 

1650

Fact_cal_sour:

 

! Address of factory calibration data source

1660

DATA 2,6,68,72,20,24,76,80,150

 

1670

!

 

 

 

1680

Fact_cal_dest

:

! Address of factory calibration data destination

1690

DATA 84,88,92,96,100,104,108,112,116

 

1700

!

 

 

 

1710

Fact_cal_len:

 

! Length of factory calibration data

 

1720

DATA 4,4,4,4,4,4,4,4,1

 

1730

!

 

 

 

1740

RESTORE Fact_cal_sour

 

1750

FOR I=1 TO 9

 

 

 

1760

READ Cal_sour_addr(I)

 

1770

NEXT I

 

 

 

1780

!

 

 

 

1790

RESTORE Fact_cal_dest

 

1800

FOR I=1 T0 9

 

 

 

1810

READ Cal_dest_addr(I)

 

1820

NEXT I

 

 

 

1830

!

 

 

 

1840

RESTORE Fact_cal_len

 

1850

FOR I=1 T0 9

 

 

 

1860

READ Cal_length(I)

 

1870

NEXT I

 

 

 

1880

!

 

 

 

1890

FOR I=1 T0 9

 

 

! Locations of good data

1900

OUTPUT @Ps;"DIAG:EEPR? ";Cal_sour_addr(I);",";Cal_length(I) ! Read good data

1910

ENTER @Ps;Cal_data$

! Enter good data

1920

OUTPUT @Ps;"DIAG:EEPR";Cal_dest_addr(I);",";Cal_length(I);”,”;Cal_data$

! Write good data to factory preset locations

 

1930

NEXT I

 

 

 

1940

!

 

 

 

1950

!

 

 

 

1960

Cal_off

 

 

 

1970

CLEAR SCREEN

 

 

1980

OUTPUT @Ps;"CaL:STATE OFF"

! Turn off cal mode

1990

!

 

 

 

2000

GOSUB Ps_error

 

! Check for errors

Figure 3-17. Initialization and Factory Preset Replacement Program Listing (Sheet 4 of 5)