Operation Verification Tests

Table 2-3 lists the requirements for operation verification, which is a subset of the performance tests.

Table 2-3. Operation Verification Tests

 

Test

Refer To

 

 

 

1

Turn-On Checkout

Power Supply Operating Manual

2

Voltage Programming and Readback Accuracy

Table 2-4

3

Current Programming and Readback Accuracy

Table 2-5

 

 

 

Note: Record the results of Tests 2 and 3 in the appropriate Performance Test Record sheets

Performance Tests

Performance tests check all the specifications of the power supply. The tests are grouped into constant-voltage mode tests (Table 2-4) and constant-current mode tests (Table 2-5).

Constant Voltage (CV) Tests

Test Setup

Connect your dc voltmeter leads to only +S and -S (see Figure 2-1), because the power supply regulates the voltage between these points, not between the + and - output terminals .

Test Procedures

Perform the test procedures in Table 2-4. The CV tests are:

Voltage Programming and Readback Accuracy

CV Load Effect

CV Source Effect

CV Noise (PARD)

Transient Recovery Time

Note

The tests are independent and may be performed in any order.

14 Verification

Page 13
Image 13
Agilent Technologies 6681A, 6684A, 6680A, 6683A Operation Verification Tests, Performance Tests, Constant Voltage CV Tests