Table 6-3. Test Points

 

TEST POINT No. & Loc.

 

 

Signal Tested

 

 

Measurement and Conditions

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

A2 GPIB BOARD

 

 

 

 

 

 

 

 

 

 

 

 

 

 

J106-4

 

Primary/chassis ground

 

 

 

Connect meter or scope common here. Then

 

 

 

 

 

 

 

 

 

 

make measurements at test points

 

 

 

 

 

 

 

 

 

 

through

.

 

 

U101- 1

 

+ 5V primary bias

 

 

 

+ 5V + 0.2V

 

 

 

U101-6

 

PCLR

 

 

 

 

 

Goes high for approximately 40 ms at power on,

 

 

 

 

PCLR*

 

 

 

 

 

then goes low.

 

 

U101-8

 

 

 

 

 

 

Held low for approximately 40 ms at power on,

 

 

 

 

STX

 

 

 

 

 

then goes high.

 

 

U110-3

 

 

 

 

 

 

Primary transmit to secondary serial data line.

 

 

 

 

SRX

 

 

 

 

 

Toggles between 0 and +5V.

 

 

U111-6

 

 

 

 

 

 

Primary receive from secondary serial data line.

 

 

 

 

FPRX

 

 

 

 

 

Toggles between 0 and +5V.

 

 

U119-4

 

 

 

 

 

 

Primary receive from front panel serial data

 

 

 

 

FPTX

 

 

 

 

 

line. Toggles between 0 and +5V.

 

 

U119-18

 

 

 

 

 

 

Primary transmit to front panel serial data line.

 

 

 

 

 

 

 

 

 

 

Toggles between 0 and +5V.

 

 

 

 

 

 

AC Input Board

 

 

 

 

 

 

 

 

 

 

 

 

 

 

Connect meter between TP

and

, or

and

, or

and

, and measure approximately 220VAC or

 

 

 

 

440VAC. Measurements at test points Rthrough X.

 

 

 

J418-1/J417-7

 

AC Mains Voltage

 

 

 

 

 

 

 

 

AC Mains Voltage

 

 

 

 

 

 

 

J418-2/J417-8

 

 

 

 

 

 

 

 

J418-3/J417-9

 

AC Mains Voltage

 

 

 

 

 

 

 

K402 coil

 

24VDC Coil Voltage

 

 

 

Turns on PS, and after power-on initialization

 

 

 

 

 

 

 

 

 

 

test, reads approximately 24VDC.

 

 

 

 

 

 

DC Rail Board

 

 

 

 

 

 

J430-1 (+) to J430-2 (-)

 

Rail #1 Voltage

 

 

 

 

Measure approximately 290VDC

 

 

J431-1 (+) to J431-2 (-)

 

Rail #2 Voltage

 

 

 

 

Measure approximately 290VDC

 

 

 

 

 

 

Bias Board

 

 

 

 

 

 

J801-4 to J801-5

 

GPIB 5V Bias

 

 

 

 

+15 ± 0.9V measure approx. 11VAC

 

 

J801-1 (+) to J801-2 (-)

 

Fan voltage

 

 

 

 

Measure 24VDC

 

 

J827-1 (+) to J827-2 (-)

 

Inrush relay coil

 

 

 

Measure 24VDC

 

 

J827-1 (+) to J827-3 (-)

 

Rail relay coil

 

 

 

 

Measure 24VDC after power-on initialization

 

 

 

 

 

 

 

 

 

 

test.

 

 

 

 

 

 

 

 

 

 

 

 

 

 

R331

 

24V Primary

 

 

 

 

Measure approximately 34VAC

 

 

R330

 

-15Vs secondary bias

 

 

 

-15 ± 0.9V

 

 

 

+C354

 

-25V

 

 

 

 

 

-22.5V to -27.5V

 

 

U310-16

 

RELAY ON*

 

 

 

 

0V

 

 

 

 

 

 

 

 

 

 

 

U308-6

 

RESET

 

 

 

 

 

0V

 

 

 

 

 

 

 

 

 

 

 

 

U308-7

 

BIAS OK

 

 

 

 

+5V

 

 

 

 

 

 

 

 

 

Diagrams 117