Figure 3-16. Signature Analysis Connections, (Sheet 2 of 2)

Table 3-6. Front Panel SA Test

Description: These signatures check front panel microprocessor AlU3.

Valid A1U4 ROM Firmware Revision: A.01.07

Test Setup: See Figure 3-17.

1 .Turn off the power supply and remove the top cover.

2.To gain access to A1 Front Panel Board, perform steps 1 and 2 of the disassembly procedure for A1 Front Panel Assembly (see "Disassembly Procedures").

3.Connect SA jumper of connector J3 on A1 Front Panel Board (see Figure 3-15).

4.Connect signature analyzer CLOCK, START, STOP, and GROUND inputs and setup as shown in Figure 3-16.

5.Turn on the power supply and use the signature analyzer probe to take the following signatures:

Power:

5V = 3395

 

Microprocessor:

AlU3-15 = 3395

A1U3-29 = 1029

 

AlU3-19 = 552U

AlU3-30 = 0295

 

AlU3-20 = 954C

AlU3-31 = 0000

 

AlU3-21 = A552

AlU3-32= 3395

 

AlU3-22 = 2954

AlU3-33 = 0008

 

AlU3-23 = 0A55

AlU3-34 = 040A

 

AlU3-24 = 3395

AlU3-35 = 0102

 

AlU3-25

= 3395

AlU3-38 = 0002

 

AlU3-26

= 0000

AlU3-39 = 0020

 

AlU3-27

= 0000

AlU3-42 = 3395

 

AlU3-28

= 40A5

 

6.After completing the tests, be sure to return the J3 jumper to its original position.

60 Troubleshooting

Page 59
Image 59
Agilent Technologies 6682A, 6684A, 6680A, 6683A, 6681A service manual Valid A1U4 ROM Firmware Revision A.01.07