Agilent Technologies 6684A, 6680A Constant Voltage CV Tests Action Normal Result, CV Load Effect

Models: 6682A 6680A 6681A 6683A 6684A

1 156
Download 156 pages 32.3 Kb
Page 15
Image 15

Table 2-4. Constant Voltage (CV) Tests

 

Action

Normal Result

 

 

 

Voltage Programming and Readback Accuracy

This test verifies that the voltage programming, GPIB readback (GPIB system power supplies only), and front panel display functions are within specifications. With system power supplies, values read back over the GPIB should be the same as those displayed on the front panel.

1

2

3

4

5

Turn off the power supply and connect a DVM across +S and -S (see Fig. 2-1).

Turn on the power supply with no load and program the output for 0 volts and maximum programmable current (see Table 2-2).

Record voltage readings at DVM and on front panel display.

Program voltage to full scale (see Table 2-2).

Record voltage readings of DVM and on front panel display.

CV annunciator on. Output current near 0.

Readings within specified Low Voltage limits.

Readings within specified High Voltage limits.

CV Load Effect

This test measures the change in output voltage resulting from a change in output current from full-load to no-load.

1

2

3

4

5

6

Turn off the power supply and connect a DVM across +S and -S(see Fig. 2-1).

Turn on the power supply and program the current to its maximum programmable value and the voltage to its full-scale value (see Table 2-2).

Adjust the load to produce full-scale current (see Table 2-2) as shown on the front panel display.

Record voltage reading of the DVM.

Adjust load to draw 0 amperes (open load). Record voltage reading of the DVM.

Check test result.

CV annunciator is on. If it is not, adjust the load to slightly reduce the output current until the annunciator comes on.

The difference between the DVM readings in steps 4 and 5 are within the specified Load Effect limits.

CV Source Effect

This test measures the change in output voltage resulting from a change in ac line voltage from its minimum to maximum value within the line voltage specifications.

1

Turn off the power supply and connect the ac power input through a variable-voltage transformer.

16 Verification

Page 15
Image 15
Agilent Technologies 6684A Constant Voltage CV Tests Action Normal Result, Voltage Programming and Readback Accuracy