43
Programming Reference
SCPI Instrument Command List
Command :ARM:SOUR
Long :ARM[:SEQuence[1]|STARt][:LAYer]:SOURce
Form Set & Query
Parameter IMMediate|INTernal[1]|INTernal2|EXTernal[1]|MANual
*RST value IMM
Description Use this command to select the triggering mode of the instrument by
selecting the source of the arming signal:
Use :ARM:SENSe EDGE | LEVel to choose between Triggered and
Gated.
Command :DISP
Long :DISPlay[:WINDow][:STATe]
Form Set & Query
Parameter ON | OFF | 1 | 0
*RST value ON
Description Use this command to turn the front panel display on and off. Switching
off the display improves the programming speed of the instrument.
*RST switches the display ba ck on. Use :SYSTem:PRESet to perform an
*RST without switching the display back on.
Example To switch off the front panel display:
DISP OFF
Triggering Source :ARM:SOURce Mode
Internal Osc.
PLL
EXT INPUT
MAN key
IMMediate|INTernal[1]
INTernal2
EXTernal1
MANual
Continuous
Triggered| Gated by PLL
Triggered| Gated by EXT IN
Triggered| Gated by MANKey