Programming Reference

SCPI Instrument Command List

Command

:SYST:WARN:STR?

Long

:SYSTem:WARNing:STRing?

Form

Query

*RST value

Not applicable

Description

Use this command to read all the currently active warning messages. The

 

warning messages are concatenated to form a single string with a “;” as

 

separator between the messages.

Command

:SYST:WARN:BUFF?

Long

:SYSTem:WARNing:BUFFer?

Form

Query

*RST value

Not applicable

Description

Use this command to read the maximum possible number of characters

 

that could be returned by :SYST:WARN:STR? if all warnings were active.

Command

:TRIG:COUNt

 

Long

:TRIGger[:SEQuence[1]]:COUNt

Form

Set & Query

 

Parameter

Numeric

 

*RST value

1

 

Specified limits

1 to 65 536

 

Description

Use this command to set/read the number of trigger events (pulse

 

periods) to be generated for each arming event. This corresponds to

 

selecting the event mode on the MODE/TRIGGER SCREEN:

 

PULSES

Set a trigger count to 1 so that a single pulse period is generated for

 

 

each arming event. The instrument is in pulse (stream) mode.

 

BURST of

Set a trigger count of 2 to 65536 so that a burst of 2 to 65536 pulse

 

 

periods is generated for each arming event.

84

Page 84
Image 84
Agilent Technologies 81101A manual Systwarnstr?, Separator between the messages, Systwarnbuff?, TRIGCOUNt

81101A specifications

Agilent Technologies 81101A is a high-performance pulse generator designed to meet the rigorous demands of today's electronic testing environments. This versatile instrument is crucial for engineers and researchers who require precise timing characteristics and reliability for their testing applications.

One of the standout features of the 81101A is its ability to generate fast, precise pulses with rise and fall times of less than 1 ns. This capability makes it ideal for testing high-speed digital circuits, as it can accurately simulate signal behavior in real-world conditions. The generator operates across a frequency range of DC to 1 GHz, allowing it to cater to a variety of applications, including telecommunications, semiconductor testing, and signal integrity investigations.

The 81101A is equipped with advanced modulation capabilities, including amplitude modulation, frequency modulation, and pulse-width modulation. These features enable users to create complex waveforms that can mimic real-world signals. Additionally, the generator supports dual-channel operation, allowing simultaneous pulse generation on two outputs, which is particularly useful for testing differential signaling schemes.

Another significant characteristic of the 81101A is its comprehensive range of output formats. It supports single-ended and differential outputs, which can be essential for measuring the performance of various devices under different operating conditions. The instrument also offers a wide range of output amplitudes, from a few millivolts to several volts, making it flexible enough to cater to a variety of testing requirements.

The user interface of the 81101A is intuitive, featuring a large graphical display that simplifies waveform configuration and real-time monitoring. This interface allows engineers to easily set parameters and visualize pulse characteristics, reducing the time required for setup and configuration.

In terms of connectivity, the 81101A is designed to integrate seamlessly into existing test environments. It includes standard interfaces such as GPIB, USB, and LAN, facilitating automated testing procedures and communication with other laboratory equipment.

Overall, the Agilent Technologies 81101A pulse generator is an essential tool for engineers seeking precise and versatile signal generation capabilities. With its high-frequency performance, advanced modulation options, and user-friendly features, it remains a valuable asset in electronic testing and research applications.