70
Programming Reference
SCPI Instrument Command List
Command :VOLT[1]
Long [:SOURce]:VOLTage[1][:LEVel][:IMMediate][:AMPLitude]
Form Set & Query
Parameter Numeric
Parameter suffix V with engineering prefixes.
Value coupling
Range coupling With Offset, see page 71
*RST value 1.00 V
Specified limits 100 mVpp to 10.0 Vpp (values ar e valid from 50 into 50 )
Description Use this command to program the amplitude voltage of the output signal.
Note that to set the output levels in terms of voltage, you first have to
execute the [:SOURce]:HOLD VOLTage c ommand to enable the
[:SOURce]:VOLTage subsystem.
The available voltage range is limited by the combination of:
Specified current limits
Actual output impedance setting :OUTPut:IMPedance
Actual expected load impedance setting:
:OUTput:IMPedance:EXTernal
Example To set the amplitude voltage:

High = Offset + Amplitude

2

Low = Offset

2
Amplitude
:HOLD VOLT
:VOLT 5V
Enable VOLTAGE subsystem
Set OUTPUT amplitude to 5 V