Barco Fusion 4MP Changing Display Functions, Information menu, Concepts, Display information

Models: Fusion 4MP

1 35
Download 35 pages 54.24 Kb
Page 26
Image 26
Information menu

Display information

Information menu

Information

Name

Description

 

 

Product

The display type

 

 

Serial No

Indicates the display serial number

 

 

SW Number

Displays the internal software number

 

 

SW Version

Displays the current internal software version

 

 

Display Lifetime

Indicates the total time the display has been

 

operating, including the time in stand-by

 

 

Backlight Lifetime

Indicates the total time the display has been

 

operating, excluding the time in stand-by

 

 

Changing Display Functions

Concepts

The Display Function menu lets you select a display function (DF).

When you select the DICOM DF you can define the DICOM DF to be recalculated taking an ambient light condition into account.

In this case, the recalculation can be based on an illuminance of 0 lux (“Dark Room”) or a preset value (“Preset”) determined by the selected reading room.

In the latter case you must select a reading room that corresponds to the type of room the display is installed in. E.g., if the display is installed in a room where CT scans are observed, the “CT/MR/NM” reading room is the preferred selection.

The reading room is defined by 2 parameters:

26