Compaq MTEK6000 user manual Unpacking, Initial Check OUT

Models: MTEK6000

1 81
Download 81 pages 16.91 Kb
Page 11
Image 11

MTEK6000 SERIES USER'S MANUAL

Chapter 2: Installation

UNPACKING

1.Thoroughly examine the box to verify it was not damaged in shipping. If you find damage, immediately file a claim with the shipper.

2.Carefully unpack the MTEK6000 from the shipping container. Verify that the box contains every item listed on the shipping order.

INITIAL CHECK OUT

!!! CAUTION !!!

This unit contains certain electronic components that are sensitive to electrostatic discharge (ESD); therefore, proper precautions should be taken during maintenance operations to avoid ESD. It is recommended that the operator first touch the shell of the MS connector (RS-232C port) on the left side of the unit to dissipate any accumulated static charge. Additional precautions may be taken in order to minimize the possibility of ESD, including the use of a grounding wrist strap (i.e., 3M part number 2214).

3.Examine the label on the left side of the enclosure. It indicates the serial number and pressure range for your unit. Verify that these parameters match your requirements. If they do not, please contact your sales representative.

4.Open the front door by loosening the upper and lower right hand corner screws or quick release latches of the enclosure and swinging the door out. See Fig. 1-1.

5.If the unit is battery powered, verify that jumpers JP 19, 20 and 24 are in the A to B position. Load the battery pack with fresh batteries and connect it to the corrector at position J6. Repeat with the second battery pack connecting it at position J7 (see Fig. 2-1). Go to step 7.

Figure 2- 1 Power connection and configuration

6.If the unit is powered via an external power source (UPS or SPS option), make sure that jumpers JP 19, 20 and 24 are in the B to C position and connect the DC power input to the DC input screw terminals 6 (V+) and 5 (GND) prior to power up. Refer to Fig. 2-1 for locating the connection points

WARNING

The operating voltage range from an external power source is 7-18 VDC. Do NOT exceed recommended input voltage of 18 VDC.

7.When you first apply power, the display will show the first two capital letters of the first label, followed by the value and then the units. For example, the label Corrected Volume MCF with a value of 00000000 would be displayed as CV 00000000 MCF.