24 Chapter4ControllingMonitorQuality
4-2. Checking Grayscale <Grayscale Check>
This function allows you to execute Grayscale Check and view the latest r esult. In addition, when the schedule
is set using RadiCS, the next grayscale check schedule appears.
● Executing Grayscale Check
Procedure
1. Choose <RadiCS SelfQ C> from the Adjustment menu, and press .
2. Choose <Grayscale Che ck> from <RadiCS SelfQC>, and press .
3. Select “Execute” with or .
4. Press to exit.
Grayscale Check is executed.
● Viewing Grayscale Check result
Attention
• The result ca n be viewed when “QC History” is “On”.
Procedure
1. Choose <RadiCS SelfQ C> from the Adjustment menu, and press .
2. Choose <Grayscale Che ck> from <RadiCS SelfQC>, and press .
3. Select “Result” with or .
4. Select “DICOM”, “Custom”, “CAL1”, “CAL2” or “CAL3” with or .
5. Press to exit.
Up to the ve latest Grayscale Check results (Pas sed/Failed/Canceled/ ERR) appear.
4-3. Setting On/Off for Warning and QC History
<Settings>
This function allows you to set On/Off for Warning and QC H istory when the Grayscale Check result is Failed.
Procedure
1. Choose <RadiCS SelfQ C> from the Adjustment menu, and press .
2. Choose <Sett ings> from <RadiCS SelfQC>, and press .
3. Select “Warning” or “QC H istory” with or .
4. Select “On” or “Off ” with or .
5. Press to exit.