Emerson 8800 instruction manual Clean, dry atmosphere, two 3/4-14 NPT conduit ports

Models: 8800

1 154
Download 154 pages 45.63 Kb
Page 18
Image 18

OCX 8800

Instruction Manual

IM-106-880, Rev 1.0 January 2007

Specifications

 

 

 

 

NEMA 4, IP66 with fitting and pipe on reference exhaust port to

Sensors Housing

 

clean, dry atmosphere, two 3/4-14 NPT conduit ports

Electronics Housing

NEMA 4, IP66 with fitting and pipe on reference exhaust port to

 

clean, dry atmosphere, two 3/4-14 NPT conduit ports

Electrical Noise

EN 61326-1, Class A

 

Certifications

 

 

 

 

C

US

APPROVED

 

 

 

 

Universal 100 to 240 VAC ±10%, 50 to 60 Hz, no switches or

Line Voltage

 

jumpers required, 3/4-14 NPT conduit port

Pollution Degree

2

 

 

 

II

 

 

Over Voltage Category

 

 

Relative Humidity

5 to 95% (non-condensing)

Isolated Output

 

 

 

Oxygen

4-20 mAdc, 950 ohm maximum, with HART capability

Combustibles

4-20 mAdc, 950 ohm maximum

Alarm

Alarm output relay - dry contact, form C, 30mA, 30VDC capacity

 

750 W maximum

 

 

Power Consumption

 

 

Mounting and Mounting

 

 

 

Positions

 

 

 

 

Flange

 

 

Sensor Housing

 

 

Electronics Housing

Integral to sensor housing

NOTE

All static performance characteristics are with operating variables constant. Specifications subject to change without notice.